Background And Objective: Hyperglycemia occurs commonly in patients admitted to medical intensive care units (MICUs). Whether real-time (RT) continuous glucose monitoring (CGM) improves glycemic control and variability and reduces hypoglycemia in severely ill MICU patients with an Acute Physiology and Chronic Health Evaluation II (APACHE-II) score of ≥20 has not been studied.

Subjects And Methods: Thirty-five patients (66 ± 10 years of age; APACHE-II score, 28 ± 6) were randomly assigned to RT-CGM (n = 16) using the GlucoDay(®)S (A. Menarini Diagnostics, Florence, Italy) device or to blinded CGM. Insulin was infused using a modified Yale protocol targeting a blood glucose level between 80 and 120 mg/dL. Outcome measures were percentage of time in normoglycemia (80-110 mg/dL) and in hypoglycemia (<60 mg/dL), glycemic variability (SD, coefficient of variation, mean amplitude of glucose excursions, and mean of daily differences), and CGM accuracy (error grid analyses, Bland-Altman bias plot, and mean absolute relative deviation).

Results: During 96 h of monitoring, glycemia reached target (80-110 mg/dL) in 37 ± 15%, was between 70 and 180 mg/dL in 91 ± 10%, and <60 mg/dL in 2 ± 2% of the time. In the RT-CGM group glycemia averaged 119 ± 17 mg/dL versus 122 ± 11 mg/dL in the control group. Parameters of glucose variability and percentages of time at target glycemia and in hypoglycemia were similar between groups. GlucoDayS values and arterial glycemia correlated well, with 98.6% of data falling in Zones A and B of the error grid analysis. Mean absolute relative devation was 11.2%.

Conclusions: RT-CGM did not ameliorate glucose control or variability; neither did it reduce the number of hypoglycemic events, but our insulin infusion protocol led to overall good glucose control without a significant hypoglycemia risk, making further improvement difficult.

Download full-text PDF

Source
http://dx.doi.org/10.1089/dia.2015.0151DOI Listing

Publication Analysis

Top Keywords

glycemic control
8
medical intensive
8
intensive care
8
real-time continuous
8
continuous glucose
8
glucose monitoring
8
apache-ii score
8
randomized evaluation
4
evaluation glycemic
4
control medical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!