Imaging start time of single-photon emission computed tomography (SPECT) scan is recommended between 3 hours and 6 hours after injection of ¹²³I-ioflupane in Ioflupane clinical practice guidelines. But image includes the effect of physiological actions of the human body and the attenuation at 13.27 hours physical half-life of ¹²³I. Therefore, we evaluated the effect of the image by the elapsed time of imaging start time. Optimal cut-off frequency of Butterworth filter were examined phantom by normalized mean square error. Count was reduced by 23.1% in 5 hours, but cut-off frequency of Butterworth filter was 0.11 cycle/pixel in 0-5 hours from phantom data. Ten subjects (age 55-85 years) were injected with ¹²³I-ioflupane of 110-199 MBq into the vein only once. And we examined the specific binding ratio (SBR) in inspection of injection after 5. 5 hours and 3 hours. Decrease of counts value increased the coefficient of variance (CV) between 3 hours and 5.5 hours after injection, but the improvement of statistical noise by pre-processing filter reduced the CV. No significant difference in the result of SBR was found between 3 hours and 5. 5 hours after injection. Our results suggest that imaging start time of SPECT scan is recommended between 3 hours and 5.5 hours after injection of ¹²³I-ioflupane.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.6009/jjrt.2015_JSRT_71.8.670 | DOI Listing |