Background: Robotic assistance for the placement of pedicle screws has been established as a safe technique. Nonetheless rare instances of screw misplacement have been reported.The aim of the present retrospective study is to assess whether experience and time affect the accuracy of screws placed with the help of the SpineAssist™ robot system.

Methods: Postoperative computed tomography (CT) scans of 258 patients requiring thoracolumbar pedicle screw instrumentation from 2008 to 2013 were reviewed. Overall, 13 surgeons performed the surgeries. A pedicle breach of >3 mm was graded as a misplacement. Surgeons were dichotomised into an early and experienced period in increments of five surgeries.

Results: In 258 surgeries, 1,265 pedicle screws were placed with the aid of the robot system. Overall, 1,217 screws (96.2 %) were graded as acceptable. When displayed by surgeon, the development of percent misplacement rates peaked between 5 and 25 surgeries in 12 of 13 surgeons. The overall misplacement rate in the first five surgeries was 2.4 % (6/245). The misplacement rate rose to 6.3 % between 11 and 15 surgeries (10/158; p = 0.20), and reached a significant peak between 16 and 20 surgeries with a rate of 7.1 % (8/112; p = 0.03). Afterwards, misplacement rates declined.

Conclusions: A major peak in screw inaccuracies occurred between cases 10 and 20, and a second, smaller one at about 40 surgeries. One potential explanation could be a transition from decreased supervision (unskilled but aware) to increased confidence of a surgeon (unskilled but unaware) who adopts this new technique prior to mastering it (skilled). We therefore advocate ensuring competent supervision for new surgeons at least during the first 25 procedures of robotic spine surgery to optimise the accuracy of robot-assisted pedicle screws.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00701-015-2535-0DOI Listing

Publication Analysis

Top Keywords

pedicle screws
12
robotic spine
8
spine surgery
8
misplacement rates
8
misplacement rate
8
surgeries
7
misplacement
6
pedicle
5
screws
5
unskilled unawareness
4

Similar Publications

3D-printed (3DP) drill guides have demonstrated significant potential to accurately guide pedicle screw insertion in spinal surgery. However, their role in the upper cervical spine is not well established. This review aimed to compare the efficacy and safety of 3DP drill guides to the conventional fluoroscopic-guided free-hand technique for pedicle screw insertion in the upper cervical spine.

View Article and Find Full Text PDF

Introduction: Congenital lumbar kyphosis is present in about 15% of patients with myelomeningocele. Worsening of deformity with complications such as chronic skin ulcers and bone exposure is common. In patients under 8 years of age, treatment becomes even more challenging: in addition to resecting the apex of the kyphotic deformity, we should ideally stabilize the spine with fixation methods that do not interrupt the growth of the rib cage, associated with the challenging pelvic fixation in this population.

View Article and Find Full Text PDF

Objective: To investigate the predictive ability of the MRI-based vertebral bone quality (VBQ) score for pedicle screw loosening following instrumented transforaminal lumbar interbody fusion (TLIF).

Methods: Data from patients who have received one or two-level instrumented TLIF from February 2014 to March 2015 were retrospectively collected. Pedicle screw loosening was diagnosed when the radiolucent zone around the screw exceeded 1 mm in plain radiographs.

View Article and Find Full Text PDF

A 3D-printed guide is an effective method for accurately placing pedicle screws in dog vertebrae. While a conventional drill guide allows precise pilot hole formation, it can lead to potential screw wobbling during insertion. In this study, we applied a technique that assists with both drilling and screw insertion, and we compared the accuracy of screw placement using this approach with that achieved by the conventional drill guide.

View Article and Find Full Text PDF

Objective: Combining oblique lumbar interbody fusion (OLIF) with posterior pedicle screw fixation (PPSF) has been proposed to reduce cage subsidence, especially in osteoporotic spines. Recently, anterolateral screw-rod fixation has gained interest as it allows direct pathology observation and avoids a posterior approach. However, controversies exist between anterolateral screw fixation systems and traditional PPSF due to variations in osteoporotic vertebral mineral density, screw fixation positions, and fixation methods (bicortical vs.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!