A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1057
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3175
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Continuous intra-arterial blood glucose monitoring using quenched fluorescence sensing in intensive care patients after cardiac surgery: phase II of a product development study. | LitMetric

Purpose: Variations in blood glucose (BG), hyperglycaemia and hypoglycaemia are associated with adverse clinical outcomes in intensive care unit patients. Continuous glucose monitoring (CGM) offers the potential to improve BG control, leading to improved patient outcomes. In our product development study, we determined the safety and performance of the GluCath Intravascular CGM System for up to 48 hours in 20 patients admitted to the ICU after cardiac surgery.

Methods: The GluCath system uses a quenched chemical fluorescence mechanism to optically measure glucose in blood. After undergoing elective cardiac surgery, 20 patients had a GluCath sensor inserted through a preexisting radial artery (RA) catheter, and BG was monitored for up to 48 hours. Qualitative measures included effects on patient care, blood pressure monitoring, and ease of blood sampling through the arterial catheter. Safety assessment of the sensor involved ultrasound (US) monitoring for intra-arterial thrombi. Quantitative measures were the accuracy of the sensor in comparison with the reference analyser, and the proportion of paired BG measurements that were compliant with the ISO15197:2003 and CLSI POCT 12-A3 accuracy reference standards. BG was managed according to usual protocols.

Results: Twenty sensors were successfully deployed through pre-existing RA catheters and stayed in the RA of the 20 patients for between 6 and 48 hours, with a median time of 45.0 hours (interquartile range, 42.0-47.3 hours). Sixteen of the inserted sensors (80%) remained in situ for more than 40 hours. Three catheters were removed due to clinically significant sampling difficulty or waveform dampening. Two patients had US evidence of a thrombus; in neither patient was the sensor removed early, and there were no significant sequelae or adverse effects detected. For the 758 paired measurements available for performance analysis, reference BG values ranged between 5.3 mmol/L and 12.8 mmol/L. Of the 758 paired sensor measurements, 735 (97.0%) met the ISO15197:2003 criteria (within 20% of a reference measurement when BG is ≥ 4.2 mmol/L [75 mg/dL]), and 648 (85.5%) met the CLSI POCT 12-A3 criteria (within 12.5% of a reference measurement when the BG level is ≥ 5.6 mmol/L [100mg/dL] and within 0.67 mmol/L [12 mg/dL] when the BG level is < 5.6mmol/L [100 mg/dL]). The aggregate mean absolute relative difference (MARD) between the sensor and the reference BG was 6.4%, with individual sensor MARDs ranging from 3.6% to 12.4%.

Conclusions: The GluCath system, using quenched fluorescence sensing, was safe and showed acceptable accuracy when deployed for up to 48 hours in ICU patients after elective cardiac surgery.

Download full-text PDF

Source

Publication Analysis

Top Keywords

cardiac surgery
12
blood glucose
8
glucose monitoring
8
quenched fluorescence
8
fluorescence sensing
8
intensive care
8
product development
8
development study
8
glucath system
8
system quenched
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!