Background And Purpose: The latency of a multileaf collimator (MLC) tracking system used to overcome respiratory motion causes misalignment of the treatment beam with respect to the gross tumour volume, which may result in reduced target coverage. This study investigates the magnitude of this effect.

Material And Methods: Simulated superior-inferior breathing motion was used to construct histograms of isocentre offset with respect to the gross tumour volume (GTV) for a variety of tracking latencies. Dose distributions for conformal volumetric modulated arc therapy (VMAT) arcs were then calculated at a range of offsets and summed according to these displacement histograms. The results were verified by delivering the plans to a Delta(4) phantom on a motion platform.

Results: In the absence of an internal target margin, a tracking latency of 150 ms reduces the GTV D95% by approximately 2%. With a margin of 2 mm, the same drop in dose occurs for a tracking latency of 450 ms. Lung V(13Gy) is unaffected by a range of latencies. These results are supported by the phantom measurements.

Conclusions: Assuming that internal motion can be modelled by a rigid translation of the patient, MLC tracking of conformal VMAT can be effectively accomplished in the absence of an internal target margin for substantial breathing motion (4 s period and 20 mm peak-peak amplitude) so long as the system latency is less than 150 ms.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4728182PMC
http://dx.doi.org/10.1016/j.radonc.2015.07.044DOI Listing

Publication Analysis

Top Keywords

mlc tracking
12
tracking latency
12
conformal volumetric
8
volumetric modulated
8
modulated arc
8
arc therapy
8
therapy vmat
8
respect gross
8
gross tumour
8
tumour volume
8

Similar Publications

Development and validation of the Normalized Organoid Growth Rate (NOGR) metric in brightfield imaging-based assays.

Commun Biol

December 2024

Center for Oncological Research (CORE), Integrated Personalized & Precision Oncology Network (IPPON), University of Antwerp, Wilrijk, Belgium.

This study focuses on refining growth-rate-based drug response metrics for patient-derived tumor organoid screening using brightfield live-cell imaging. Traditional metrics like Normalized Growth Rate Inhibition (GR) and Normalized Drug Response (NDR) have been used to assess organoid responses to anticancer treatments but face limitations in accurately quantifying cytostatic and cytotoxic effects across varying growth rates. Here, we introduce the Normalized Organoid Growth Rate (NOGR) metric, specifically developed for brightfield imaging-based assays.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the effectiveness of using 3D-spheroids of human adipose stem cells (hASCs) for treating fecal incontinence by focusing on their ability to develop into smooth muscle cells (SMCs) when implanted in a mouse model.
  • Researchers created a targeted model by freezing part of the internal anal sphincter (IAS) in rats and then analyzed SMC markers and related proteins after implanting either 3D-spheroids or 2D-cultured cells.
  • Results showed that 3D-spheroids had greater expression of SMC markers and induced better contractile characteristics over time compared to 2D-hASCs, indicating a potential therapeutic benefit for fecal incontinence treatment.
View Article and Find Full Text PDF

Background: Canada has one of the highest incidences of colorectal cancer (CRC) worldwide. CRC screening improves CRC outcomes and is cost-effective. This study compares Canadian CRC screening programs using essential elements of an organized screening program outlined by the International Agency for Research on Cancer (IARC).

View Article and Find Full Text PDF

Purpose: This study presents a novel and comprehensive framework for evaluating magnetic resonance guided radiotherapy (MRgRT) workflow by integrating the Failure Modes and Effects Analysis (FMEA) approach with Time-Driven Activity-Based Costing (TDABC). We assess the workflow for safety, quality, and economic implications, providing a holistic understanding of the MRgRT implementation. The aim is to offer valuable insights to healthcare practitioners and administrators, facilitating informed decision-making regarding the 0.

View Article and Find Full Text PDF

Background: Several (online) adaptive radiotherapy procedures are available to maximize healthy tissue sparing in the presence of inter/intrafractional motion during stereotactic body radiotherapy (SBRT) on an MR-linac. The increased treatment complexity and the motion-delivery interplay during these treatments require MR-compatible motion phantoms with time-resolved dosimeters to validate end-to-end workflows. This is not possible with currently available phantoms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!