Background: Vitamin C is a strong reducing agent found at high levels in various foods, and it may influence the results of urine strip tests even at an ordinary consumption levels. After oral administration, we measured urine vitamin C levels using urine strips and evaluated whether vitamin C interfered with various test items. The utility of a urine strip with a vitamin C indicator was assessed.

Methods: Thirty-three healthy volunteers each ingested 1,000 mg of vitamin C. Their urine samples were tested for vitamin C using a URiSCAN 11 strip (YD Diagnostics, Korea) before and after administration of vitamin C. Standard materials were added to normal pooled urine to generate urine samples with various concentrations of the analytes tested (blood, bilirubin, nitrite, leukocytes, and glucose), and vitamin C was spiked to predetermined levels. These samples were then tested using two urine strips - URiSCAN and Chemstrip test strip (Roche Diagnostics, Germany) - to evaluate interference from vitamin C. In clinical samples with positive vitamin C results, microscopic and chemical analyses were also conducted to examine the differences.

Results: Thirteen urine samples from the 33 volunteers were positive for vitamin C before ingestion, and all subjects were positive after ingestion. Vitamin C spiking of urine demonstrated false-negative results at various concentrations. Of 159 specimens with positive results for vitamin C, 14 showed discrepant results after additional confirmatory tests.

Conclusions: Vitamin C in urine can cause significant interference with urine strip tests. A urine strip with a vitamin C indicator is useful to reduce the risk of incorrect results in regard to disease states.

Download full-text PDF

Source

Publication Analysis

Top Keywords

urine strip
16
vitamin
15
urine
14
vitamin urine
12
urine samples
12
positive vitamin
12
strip tests
8
urine strips
8
strip vitamin
8
vitamin indicator
8

Similar Publications

Current American Urological Association guidelines recommend that patients with acute obstructive kidney stone requiring continuous anticoagulation/antiplatelet therapy should not be treated by shockwave lithotripsy or percutaneous nephrolithotomy because of the risk of catastrophic renal hemorrhage possible with those techniques. Currently, ureteroscopy is the only recommended surgical treatment. We evaluated if burst wave lithotripsy (BWL) could be used in these cases by treating pigs with BWL while undergoing anticoagulation therapy.

View Article and Find Full Text PDF

Urinalysis is one of the predominant tools for clinical testing owing to the abundant composition, sufficient volume, and non-invasive acquisition of urine. As a critical component of routine urinalysis, urine protein testing measures the levels and types of proteins, enabling the early diagnosis of diseases. Traditional methods require three separate steps including strip testing, protein/creatinine ratio measurement, and electrophoresis respectively to achieve qualitative, quantitative, and classification analyses of proteins in urine with long time and cumbersome operations.

View Article and Find Full Text PDF

Background: One method for noninvasive and simple urinary microalbumin testing is urine test strips. However, when visually assessing urine test strips, accurate assessment may be difficult due to environmental influences-such as lighting color and intensity-and the physical and psychological influences of the assessor. These complicate the formation of an objective assessment.

View Article and Find Full Text PDF

Background: Urogenital schistosomiasis is a persistent public health problem in many rural areas of Yemen. Since 2014, epidemiology has not been assessed in Amran governorate, north of Yemen, where is known to be highly endemic. Therefore, this study determined the prevalence and risk factors associated with infection among schoolchildren in Kharif district of the governorate.

View Article and Find Full Text PDF

Background: The rapid and sensitive detection of nitrite is important to human health protection due to its carcinogenic and teratogenic risks with excessive intake. The Griess assay is widely applied for the design of nitrite detection system. However, its relatively slow reaction kinetics and sole colorimetry mode might limit it's the sensitivity and practical application.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!