Background: The 12-lead electrocardiographic screening for the prevention of sudden cardiac death in young competitive athletes is not cost-effective and thus not routinely recommended. We investigate whether a less expensive wireless electrocardiographic transmission device can be used to screen for the prevention of sudden cardiac death in this population.
Methods: During pre-participation screening, twenty college football players underwent two electrocardiograms: a conventional 12-lead electrocardiogram and a wireless 9-lead electrocardiogram. We compared several electrocardiographic parameters (QRS duration, left ventricular hypertrophy using the Cornell voltage criteria and the Sokolow-Lyon criteria, ST deviation and corrected QT interval) to determine the correlation.
Results: The QRS duration, left ventricular hypertrophy using the Cornell voltage criteria and the Sokolow-Lyon criteria and corrected QT interval exhibited significant correlation between the two types of electrocardiograms (correlation coefficient 0.878, 0.630, 0.770 and 0.847, respectively with P values of 0.01, 0.003, 0.01 and 0.01, respectively). ST deviation in V1 was weakly correlated between the two types of electrocardiograms without statistical significance (correlation coefficient 0.360 with a P value of 0.119).
Conclusions: Our newly developed wireless 9-lead electrocardiogram demonstrated significant correlations with a conventional 12-lead electrocardiogram in terms of QRS duration, left ventricular hypertrophy and corrected QT interval.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4531846 | PMC |
http://dx.doi.org/10.1186/s13104-015-1311-9 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!