In this paper, we introduce a high throughput LCMS/UV/CAD/CLND system that improves upon previously reported systems by increasing both the quantitation accuracy and the range of compounds amenable to testing, in particular, low molecular weight "fragment" compounds. This system consists of a charged aerosol detector (CAD) and chemiluminescent nitrogen detector (CLND) added to a LCMS/UV system. Our results show that the addition of CAD and CLND to LCMS/UV is more reliable for concentration determination for a wider range of compounds than either detector alone. Our setup also allows for the parallel analysis of each sample by all four detectors and so does not significantly increase run time per sample.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2015.07.112DOI Listing

Publication Analysis

Top Keywords

charged aerosol
8
range compounds
8
clnd lcms/uv
8
aerosol detector/chemiluminescent
4
detector/chemiluminescent nitrogen
4
nitrogen detector/liquid
4
detector/liquid chromatography/mass
4
chromatography/mass spectrometry
4
system
4
spectrometry system
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!