Objective: To investigate the influence of the number of T-junctions between central venous catheter and pressure transducer on measurement of central venous pressure ( CVP ) in patients.

Methods: A prospective controlled study was conducted. The patients with CVP monitoring in Department of Critical Care Medicine of the Fifth Center Hospital in Tianjin from February to October in 2014 were enrolled. The patients were divided into three groups according to the number of T-junction between central venous catheter and pressure transducer: without T-junction control group and 1, 2, 3 T-junctions groups. In each patient, corresponding CVP values with different number of T-junctions placed between the central venous catheter and pressure sensors were determined within a certain period, and a square-wave graphic was obtained and preserved on the monitor. The own frequency ( fn ) and the attenuation coefficient ( D ) of the system of pressure measurement were calculated after measurement of the shock wave following a square-wave to obtain the distance between two vibrations and the amplitude of the shock wave. The difference in CVP, fn and D were compared among the groups.

Results: A total of 20 cases were enrolled, and 150 groups of data were collected. (1) With the increase in the number of T-junction, CVP showed a tendency of gradual reduction. The CVP of the groups of control and 1, 2, 3 T-junctions was ( 7.00±1.60 ), ( 7.00±3.00 ), ( 5.00±2.00 ), and ( 4.00±1.00 ) mmHg ( 1 mmHg = 0.133 kPa ), respectively. The CVP of 3 T-junctions group was significantly lower than that of the control group ( F = 9.333, P = 0.015 ). (2) With an increase in the number of T-junction, fn showed a tendency of gradual increase. The fn of groups control and 1, 2, 3 T-junctions was ( 12.30±0.79 ), ( 16.00±0.91 ), ( 18.10±1.75 ), ( 20.90±2.69 ) Hz, respectively. The fn of 1, 2, 3 T-junctions group was significantly higher than that of the control group ( F1 = 45.962, F2 = 45.414, F3 = 46.830, all P = 0.000 ); the fn of groups 2 and 3 T-junctions was significantly higher than that of 1 T-junction group ( F1= 5.827, P1= 0.042; F2 = 15.038, P2 = 0.004 ), but there was no significant difference between the groups of 2 T-junctions and 3 T-junctions ( F = 3.800, P = 0.087 ). (3) With an increase of the number of T-junction, D also showed a tendency of gradual increase. The D of 1, 2, 3 T-junction group was 1.62±0.27, 1.60±0.22, 1.82±0.25, and 2.15±0.58, respectively. There were no differences among four groups.

Conclusions: After the application of T-junctions between central venous catheter and pressure transducer, CVP values will be underestimated, the reason of which is considered to be the increase in length and thinner lumen of the T-junctions.

Download full-text PDF

Source
http://dx.doi.org/10.3760/cma.j.issn.2095-4352.2015.08.015DOI Listing

Publication Analysis

Top Keywords

central venous
28
venous catheter
20
catheter pressure
20
pressure transducer
16
t-junctions central
16
number t-junction
16
t-junctions
13
control group
12
increase number
12
tendency gradual
12

Similar Publications

Introduction: Haemodynamic atrioventricular delay (AVD) optimisation has primarily focussed on signals that are not easy to acquire from a pacing system itself, such as invasive left ventricular catheterisation or arterial blood pressure (ABP). In this study, standard clinical central venous pressure (CVP) signals are tested as a potential alternative.

Methods: Sixteen patients with a temporary pacemaker after cardiac surgery were studied.

View Article and Find Full Text PDF

Background: The cannulation of the internal jugular vein (IJV) is a frequent procedure in critically ill patients. According to the guidelines, real-time ultrasound navigation is recommended. Traditional techniques pose several disadvantages, such as suboptimal needle visualization.

View Article and Find Full Text PDF

Purpose Of Review: Critical Care Echocardiography (CCE) is now established as an important tool in the intensive care unit (ICU). This paper aims to examine the expanding role of cardiovascular ultrasound in the ICU, focusing on its applications, benefits, and challenges, while highlighting recent advancements shaping the future of critical care echocardiography.

Recent Findings: Non-invasive echocardiographic measurement of hemodynamic parameters including stroke volume, cardiac output, left ventricular filling pressures, and pulmonary pressures have been well-validated against invasive measurements.

View Article and Find Full Text PDF

Pulmonary cavernous hemangioma: a case report.

BMC Pulm Med

January 2025

Department of Medical Imaging, Baoji Central Hospital, Baoji, China.

Background: Cavernous hemangiomas can occur in various internal organs like the liver, kidney, bladder, and skin, or even in subcutaneous tissues. However, they rarely occur in the lungs, making pulmonary cavernous hemangiomas (PCH) an uncommon finding. Herein, we report a rare case of pulmonary cavernous hemangioma that was surgically resected.

View Article and Find Full Text PDF

Introduction: In many countries, hemodialysis is still the most common method of renal replacement. Significant regional variations exist in creating and maintaining vascular access for hemodialysis. Therefore, analyzing national registries with vascular access data is essential for developing effective national treatment strategies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!