AI Article Synopsis

  • Animal studies are essential for validating the safety and effectiveness of a new ultrasound-based battery recharging system called USER™.
  • Laboratory tests showed that the system could deliver power effectively through various tissue depths, achieving up to 600 mW through 15 mm and 50 mW through 50 mm of tissue.
  • In an extensive five-week study involving pigs, the device demonstrated reliable power delivery without damaging surrounding tissue, suggesting that ultrasound could safely recharge batteries in medical implants.

Article Abstract

Animal studies are an important step in proving the utility and safety of an ultrasound based implanted battery recharging system. To this end an Ultrasound Electrical Recharging System (USER™) was developed and tested. Experiments in vitro demonstrated power deliveries at the battery of up to 600 mW through 10-15 mm of tissue, 50 mW of power available at tissue depths of up to 50 mm, and the feasibility of using transducers bonded to titanium as used in medical implants. Acute in vivo studies in a porcine model were used to test reliability of power delivery, temperature excursions, and cooling techniques. The culminating five-week survival study involved repeated battery charging, a total of 10.5h of ultrasound exposure of the intervening living tissue, with an average RF input to electrical charging efficiency of 20%. This study was potentially the first long term cumulative living-tissue exposure using transcutaneous ultrasound power transmission to an implanted receiver in situ. Histology of the exposed tissue showed changes attributable primarily due to surgical implantation of the prototype device, and no damage due to the ultrasound exposure. The in vivo results are indicative of the potential safe delivery of ultrasound energy for a defined set of source conditions for charging batteries within implants.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4586961PMC
http://dx.doi.org/10.1016/j.ultras.2015.07.012DOI Listing

Publication Analysis

Top Keywords

ultrasound power
8
power delivery
8
recharging system
8
ultrasound exposure
8
ultrasound
7
power
5
vivo demonstration
4
demonstration ultrasound
4
delivery charge
4
charge implanted
4

Similar Publications

Simulating the cardiac valves is one of the most complex tasks in cardiovascular modeling. As fluid-structure interaction simulations are highly computationally demanding, machine-learning techniques can be considered a good alternative. Nevertheless, it is necessary to design many aortic valve geometries to generate a training set.

View Article and Find Full Text PDF

The Emerging Role of Sonoelastography in Pregnancy: Applications in Assessing Maternal and Fetal Health.

Diagnostics (Basel)

December 2024

Radiology and Medical Imaging Department, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Al-Kharj 16278, Saudi Arabia.

Sonoelastography, a novel ultrasound-based technique, is emerging as a valuable tool in prenatal diagnostics by quantifying tissue elasticity and stiffness in vivo. This narrative review explores the application of sonoelastography in assessing maternal and fetal health, with a focus on cervical, placental, pelvic floor, and fetal tissue evaluations. In the cervix, sonoelastography aids in predicting preterm birth and assessing labor induction success.

View Article and Find Full Text PDF

The maternal body mass index and first-trimester placental (vascular) development.

Placenta

December 2024

Department of Obstetrics and Gynecology, Erasmus MC, University Medical Center, Rotterdam, the Netherlands. Electronic address:

Background: Maternal obesity is associated with maternal complications, including hypertensive disorders of pregnancy (HDP), and related fetal complications, such as fetal growth restriction. During pregnancy, the placenta is one of the key regulators of embryonic and fetal growth. Previous studies mainly investigated placental growth by measuring postpartum placental weight.

View Article and Find Full Text PDF

Purpose: Due to the highly individualized clinical manifestation of Parkinson's disease (PD), personalized patient care may require domain-specific assessment of neurological disability. Evidence from magnetic resonance imaging (MRI) studies has proposed that heterogenous clinical manifestation corresponds to heterogeneous cortical disease burden, suggesting customized, high-resolution assessment of cortical pathology as a candidate biomarker for domain-specific assessment.

Method: Herein, we investigate the potential of the recently proposed Mosaic Approach (MAP), a normative framework for quantifying individual cortical disease burden with respect to a population-representative cohort, in predicting domain-specific clinical progression.

View Article and Find Full Text PDF

The predictive value of anti-IFI16 antibodies for the development or persistence of digital ulcers in systemic sclerosis.

Clin Rheumatol

January 2025

Department of Rheumatology, Huashan Hospital, Fudan University, No.12 Wulumuqi Zhong Road, Shanghai, 200040, China.

To evaluate the association of anti-IFI16 antibodies with peripheral vasculopathy and the predictive value of anti-IFI16 antibodies for the development or persistence of digital ulcers (DPDU) in SSc. A total of 42 SSc patients and 42 age- and sex-matched healthy controls were enrolled. Anti-IFI16 antibodies were examined by ELISA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!