Diaphragm pacing failure secondary to deteriorated chest wall mechanics: When a good diaphragm does not suffice to take a good breath in.

Respir Med Case Rep

AP-HP, Groupe Hospitalier Pitié-Salpêtrière Charles Foix, Service de Pneumologie et Réanimation Médicale (Département "R3S"), F-75013, Paris, France ; Sorbonne Universités, UPMC Univ Paris 06, UMR_S 1158 "Neurophysiologie Respiratoire Expérimentale et Clinique", F-75005, Paris, France ; INSERM, UMR_S 1158 "Neurophysiologie Respiratoire Expérimentale et Clinique", F-75005, Paris, France.

Published: August 2015

Diaphragm pacing allows certain quadriplegic patients to be weaned from mechanical ventilation. Pacing failure can result from device dysfunction, neurotransmission failure, or degraded lung mechanics (such as atelectasis). We report two cases where progressive pacing failure was attributed to deteriorated chest wall mechanics. The first patient suffered from cervical spinal cord injury at age 45, was implanted with a phrenic stimulator (intrathoracic), successfully weaned from ventilation, and permanently paced for 7 years. Pacing effectiveness then slowly declined, finally attributed to rib cage stiffening due to ankylosing spondylitis. The second patient became quadriplegic after meningitis at age 15, was implanted with a phrenic stimulator (intradiaphragmatic) and weaned. After a year hypoventilation developed without obvious cause. In relationship with complex endocrine disorders, the patient had gained 31 kg. Pacing failure was attributed to excessive mechanical inspiratory load. Rib cage mechanics abnormalities should be listed among causes of diaphragm pacing failure and it should be kept in mind that a "good diaphragm" is not sufficient to produce a "good inspiration".

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4501463PMC
http://dx.doi.org/10.1016/j.rmcr.2015.03.006DOI Listing

Publication Analysis

Top Keywords

pacing failure
20
diaphragm pacing
12
deteriorated chest
8
chest wall
8
wall mechanics
8
failure attributed
8
age implanted
8
implanted phrenic
8
phrenic stimulator
8
rib cage
8

Similar Publications

Right ventricular pacing is an effective and safe treatment option for patients experiencing symptomatic bradycardia. However, some individuals may develop left ventricular dysfunction as a consequence. Growth differentiation factor 15 (GDF-15), which is not present in a healthy adult heart, is upregulated in cardiomyocytes in response to various stress stimuli.

View Article and Find Full Text PDF

A leadless pacemaker (LP) is a modern alternative to a transvenous pacemaker, allowing certain complications to be avoided; however, some cannot be eliminated. To highlight the essential role of advanced speckle-tracking echocardiography (STE) in diagnosing pacing-induced cardiomyopathy (PICM) caused by an LP. A 79-year-old male, after LP implantation a year earlier, was admitted due to heart failure (HF).

View Article and Find Full Text PDF

Typical Atrial Flutter: A Practical Review.

J Cardiovasc Electrophysiol

January 2025

Cardiology Division, Geneva University Hospitals, Geneva, Switzerland.

Typical atrial flutter (AFL), defined as cavotricuspid isthmus (CTI)-dependent macro-re-entrant atrial tachycardia, often causes debilitating symptoms, and is associated with increased incidence of atrial fibrillation, stroke, heart failure, and death. Typical AFL occurs in patients with atrial remodeling and shares risk factors with atrial fibrillation. It is also common in patients with a history of prior heart surgery or catheter ablation.

View Article and Find Full Text PDF

Background: Conduction disturbances are a frequent occurrence after tricuspid valve surgeries, and their management is challenging.

Case Presentation: We present a case of 16-year-old male patient who presented with episodes of presyncope. At the age of 7 years, he underwent tricuspid valve replacement surgery with a biological prosthesis for infective endocarditis sourced from a gluteal abscess.

View Article and Find Full Text PDF

Transthyretin Cardiac amyloidosis (ATTR-CA) is an increasingly recognised cause of heart failure in our elderly patients with preserved ejection fraction. Patients with ATTR-CA who require permanent pacemaker implantation often have preserved ejection fraction and do not meet the clinical indication for cardiac resynchronization therapy (CRT). In these patients, left bundle branch area pacing (LBBAP) can be a reasonable option to maximise physiological activation of the left ventricle.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!