AI Article Synopsis

  • The study aimed to optimize chest compressions during out-of-hospital cardiac arrest to enhance organ perfusion and survival rates.
  • The research was conducted using a porcine model, examining relationships between mechanical measures (like compression depth and force) and physiological markers (like end-tidal CO2 and systolic blood pressure) during cardiopulmonary resuscitation (CPR).
  • Results showed moderate to strong correlations between the amplitude of the impedance cardiogram and both compression depth and other physiological parameters, suggesting that the ICG could be a valuable feedback tool for improving CPR effectiveness.

Article Abstract

Objective: Optimising the depth and rate of applied chest compressions following out of hospital cardiac arrest is crucial in maintaining end organ perfusion and improving survival. The impedance cardiogram (ICG) measured via defibrillator pads produces a characteristic waveform during chest compressions with the potential to provide feedback on cardiopulmonary resuscitation (CPR) and enhance performance. The objective of this pre-clinical study was to investigate the relationship between mechanical and physiological markers of CPR efficacy in a porcine model and examine the strength of correlation between the ICG amplitude, compression depth and end-tidal CO2 (ETCO2).

Methods: Two experiments were performed using 24 swine (12 per experiment). For experiment 1, ventricular fibrillation (VF) was induced and mechanical CPR commenced at varying thrusts (0-60 kg) for 2 min intervals. Chest compression depth was recorded using a Philips QCPR device with additional recording of invasive physiological parameters: systolic blood pressure, ETCO2, cardiac output and carotid flow. For experiment 2, VF was induced and mechanical CPR commenced at varying depths (0-5 cm) for 2 min intervals. The ICG was recorded via defibrillator pads attached to the animal's sternum and connected to a Heartsine 500 P defibrillator. ICG amplitude, chest compression depth, systolic blood pressure and ETCO2 were recorded during each cycle. In both experiments the within-animal correlation between the measured parameters was assessed using a mixed effect model.

Results: In experiment 1 moderate within-animal correlations were observed between physiological parameters and compression depth (r=0.69-0.77) and thrust (r=0.66-0.82). A moderate correlation was observed between compression depth and thrust (r=0.75). In experiment 2 a strong within-animal correlation and moderate overall correlations were observed between ICG amplitude and compression depth (r=0.89, r=0.79) and ETCO2 (r=0.85, r=0.64).

Conclusion: In this porcine model of induced cardiac arrest moderate within animal correlations were observed between mechanical and physiological markers of chest compression efficacy demonstrating the challenge in utilising a single mechanical metric to quantify chest compression efficacy. ICG amplitude demonstrated strong within animal correlations with compression depth and ETCO2 suggesting its potential utility to provide CPR feedback in the out of hospital setting to improve performance.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.resuscitation.2015.07.020DOI Listing

Publication Analysis

Top Keywords

compression depth
28
icg amplitude
16
chest compression
16
porcine model
12
cardiac arrest
12
correlations observed
12
depth
9
compression
9
impedance cardiogram
8
cardiopulmonary resuscitation
8