  • * Two patients, unable to receive heart transplants and in imminent risk of death, underwent successful C-TAH implantation, with both showing rapid recovery post-surgery.
  • * Despite initial success, one patient later died due to device failure, with autopsy findings showing no significant clot formation in the bioprosthesis.

Article Abstract

Background: The development of artificial hearts in patients with end-stage heart disease have been confronted with the major issues of thromboembolism or haemorrhage. Since valvular bioprostheses are associated with a low incidence of these complications, we decided to use bioprosthetic materials in the construction of a novel artificial heart (C-TAH). We report here the device characteristics and its first clinical applications in two patients with end-stage dilated cardiomyopathy. The aim of the study was to evaluate safety and feasibility of the CARMAT TAH for patients at imminent risk of death from biventricular heart failure and not eligible for transplant.

Methods: The C-TAH is an implantable electro-hydraulically actuated pulsatile biventricular pump. All components, batteries excepted, are embodied in a single device positioned in the pericardial sac after excision of the native ventricles. We selected patients admitted to hospital who were at imminent risk of death, having irreversible biventricular failure, and not eligible for heart transplantation, from three cardiac surgery centres in France.

Findings: The C-TAH was implanted in two male patients. Patient 1, aged 76 years, had the C-TAH implantation on Dec 18, 2013; patient 2, aged 68 years, had the implantation on Aug 5, 2014. The cardiopulmonary bypass times for C-TAH implantation were 170 min for patient 1 and 157 min for patient 2. Both patients were extubated within the first 12 postoperative hours and had a rapid recovery of their respiratory and circulatory functions as well as a normal mental status. Patient 1 presented with a tamponade on day 23 requiring re-intervention. Postoperative bleeding disorders prompted anticoagulant discontinuation. The C-TAH functioned well with a cardiac output of 4·8-5·8 L/min. On day 74, the patient died due to a device failure. Autopsy did not detect any relevant thrombus formation within the bioprosthesis nor the different organs, despite a 50-day anticoagulant-free period. Patient 2 experienced a transient period of renal failure and a pericardial effusion requiring drainage, but otherwise uneventful postoperative course. He was discharged from the hospital on day 150 after surgery with a wearable system without technical assistance. After 4 months at home, the patient suffered low cardiac output. A change of C-TAH was attempted but the patient died of multiorgan failure.

Interpretation: This preliminary experience could represent an important contribution to the development of total artificial hearts using bioprosthetic materials.

Funding: CARMAT SA.

Download full-text PDF

Source
http://dx.doi.org/10.1016/S0140-6736(15)60511-6DOI Listing

Publication Analysis

Top Keywords

patient
9
total artificial
8
artificial heart
8
artificial hearts
8
patients end-stage
8
imminent risk
8
risk death
8
failure eligible
8
patient aged
8
aged years
8

Similar Publications

The goal of this study was to determine how radiologists' rating of image quality when using 0.5T Magnetic Resonance Imaging (MRI) compares to Computed Tomography (CT) for visualization of pathology and evaluation of specific anatomic regions within the paranasal sinuses. 42 patients with clinical CT scans opted to have a 0.

View Article and Find Full Text PDF

Fluid administration is widely used to treat hypotension in patients undergoing veno-venous extracorporeal membrane oxygenation (VV-ECMO). However, excessive fluid administration may lead to fluid overload can aggravate acute respiratory distress syndrome (ARDS) and increase patient mortality, predicting fluid responsiveness is of great significance for VV-ECMO patients. This prospective single-center study was conducted in a medical intensive care unit (ICU) and finally included 51 VV-ECMO patients with ARDS in the prone position (PP).

View Article and Find Full Text PDF

Diabetes Mellitus combined with Mild Cognitive Impairment (DM-MCI) is a high incidence disease among the elderly. Patients with DM-MCI have considerably higher risk of dementia, whose daily self-care and life management (i.e.

View Article and Find Full Text PDF

Initial experience of a novel surgical assist robot "Saroa" featuring tactile feedback and a roll-clutch system in radical prostatectomy.

Sci Rep

December 2024

Department of Urology, The Jikei University School of Medicine, Kashiwa Hospital, Kashiwashita 163-1, Kashiwa, Chiba, 277-8567, Japan.

To evaluate the safety and efficacy of the Saroa Surgical Robot System in robot-assisted laparoscopic radical prostatectomy (RARP). We enrolled 60 patients who underwent RARP using either the Saroa (n = 9) or da Vinci Xi (n = 51) systems at Jikei University Kashiwa Hospital from January 2022 to March 2024. We compared preoperative characteristics, perioperative outcomes, complications, and postoperative urinary continence at three months between the two groups.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!