Evaluation of the Accuracy and Precision of a Next Generation Computer-Assisted Surgical System.

Clin Orthop Surg

Engineering and Development, Knee, Exactech Inc., Gainesville, FL, USA.

Published: June 2015

Background: Computer-assisted orthopaedic surgery (CAOS) improves accuracy and reduces outliers in total knee arthroplasty (TKA). However, during the evaluation of CAOS systems, the error generated by the guidance system (hardware and software) has been generally overlooked. Limited information is available on the accuracy and precision of specific CAOS systems with regard to intraoperative final resection measurements. The purpose of this study was to assess the accuracy and precision of a next generation CAOS system and investigate the impact of extra-articular deformity on the system-level errors generated during intraoperative resection measurement.

Methods: TKA surgeries were performed on twenty-eight artificial knee inserts with various types of extra-articular deformity (12 neutral, 12 varus, and 4 valgus). Surgical resection parameters (resection depths and alignment angles) were compared between postoperative three-dimensional (3D) scan-based measurements and intraoperative CAOS measurements. Using the 3D scan-based measurements as control, the accuracy (mean error) and precision (associated standard deviation) of the CAOS system were assessed. The impact of extra-articular deformity on the CAOS system measurement errors was also investigated.

Results: The pooled mean unsigned errors generated by the CAOS system were equal or less than 0.61 mm and 0.64° for resection depths and alignment angles, respectively. No clinically meaningful biases were found in the measurements of resection depths (< 0.5 mm) and alignment angles (< 0.5°). Extra-articular deformity did not show significant effect on the measurement errors generated by the CAOS system investigated.

Conclusions: This study presented a set of methodology and workflow to assess the system-level accuracy and precision of CAOS systems. The data demonstrated that the CAOS system investigated can offer accurate and precise intraoperative measurements of TKA resection parameters, regardless of the presence of extra-articular deformity in the knee.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4515464PMC
http://dx.doi.org/10.4055/cios.2015.7.2.225DOI Listing

Publication Analysis

Top Keywords

caos system
24
extra-articular deformity
20
accuracy precision
16
caos systems
12
errors generated
12
resection depths
12
depths alignment
12
alignment angles
12
caos
11
precision generation
8

Similar Publications

On the timescales in the chaotic dynamics of a 4D symplectic map.

Chaos

October 2024

Grupo de Caos en Sistemas Hamiltonianos, Facultad de Ciencias Astronómicas y Geofísicas, Universidad Nacional de La Plata and Instituto de Astrofísica de La Plata (CONICET), Paseo del Bosque S/N, B1900FWA La Plata, Argentina.

In this work, we investigate different timescales of chaotic dynamics in a multi-parametric 4D symplectic map. We compute the Lyapunov time and a macroscopic timescale, the instability time, for a wide range of values of the system's parameters and many different ensembles of initial conditions in resonant domains. The instability time is obtained by plain numerical simulations and by its estimates from the diffusion time, which we derive in three different ways: through a normal and an anomalous diffusion law and by the Shannon entropy, whose formulation is briefly revisited.

View Article and Find Full Text PDF

Conventional methods for detecting unsaturated fatty acids (UFAs) pose challenges for rapid analyses due to the need for complex pretreatment and expensive instruments. Here, we developed an intelligent platform for facile and low-cost analysis of UFAs by combining a smartphone-assisted colorimetric sensor array (CSA) based on MnO nanozymes with "image segmentation-feature extraction" deep learning (ISFE-DL). Density functional theory predictions were validated by doping experiments using Ag, Pd, and Pt, which enhanced the catalytic activity of the MnO nanozymes.

View Article and Find Full Text PDF

Calcium oxide adsorption of gas phase PCDD/Fs and its impact on the adsorption properties of activated carbon.

Environ Pollut

November 2024

State Key Laboratory for Clean Energy Utilization, Institute for Thermal Power Engineering, Zhejiang University, Hangzhou, 310027, China. Electronic address:

Calcium oxide (CaO), utilized in semi-dry/dry desulfurization systems at municipal solid waste incineration (MSWI) plants, demonstrates some capability to remove polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs). This study assessed the gas-phase PCDD/F removal performance of CaO, activated carbon (AC) and CaO-AC mixtures. Alone, CaO achieved removal efficiencies of only 31.

View Article and Find Full Text PDF

Automatic identification of radius and ulna bone landmarks on 3D virtual models.

Comput Biol Med

September 2024

Department of Orthopaedics and Sports Medicine, Erasmus MC, University Medical Center Rotterdam, Dr. Molewaterplein 40, Rotterdam, 3015 GD, The Netherlands.

Background: For bone morphology and biomechanics analysis, landmarks are essential to define position, orientation, and shape. These landmarks define bone and joint coordinate systems and are widely used in these research fields. Currently, no method is known for automatically identifying landmarks on virtual 3D bone models of the radius and ulna.

View Article and Find Full Text PDF

Advancements in computer-assisted orthognathic surgery: A comprehensive review and clinical application in South Korea.

J Dent

July 2024

Department of Artificial Intelligence and Robotics in Dentistry, Graduated School of Clinical Dentistry, Hallym University, Chuncheon 24252, South Korea; Institute of Clinical Dentistry, Hallym University, Chuncheon 24252, South Korea; Department of Oral and Maxillofacial Surgery, Hallym University Sacred Heart Hospital, Anyang 14066, South Korea; Dental Artificial Intelligence and Robotics R&D Center, Hallym University Medical Center, Anyang 14066, South Korea. Electronic address:

Objectives: Orthognathic surgery (OS) has evolved with technological advancements, notably through the implementation of computer-assisted orthognathic surgery (CAOS). This article aims to elucidate various types of CAOS and their efficiency and accuracy, supplemented by a thorough literature review focusing on their clinical applications in South Korea.

Study Selection, Data, And Sources: A comprehensive search strategy was employed, including systematic reviews, meta-analyses, randomized controlled trials, and observational studies published until December 2023 in the PubMed, MEDLINE, and Google Scholar databases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!