Background: We used the Medtronic Freestyle valve (Medtronic, Minneapolis, MN) as an orthotopic conduit in pulmonary valve replacement in repaired tetralogy of Fallot and as part of the Ross procedure. Midterm outcomes and hemodynamic status of this conduit were analyzed and performances in both subgroups were compared.
Methods: From February 2002 to July 2012, 115 Freestyle valves were implanted in 52 patients with tetralogy of Fallot and 63 patients within the Ross procedure. Preoperative and perioperative data were reviewed retrospectively in this bicentric study.
Results: Mean age at valve surgery was 37 ± 13 years. Median implanted valve size was 27 mm (21 to 29). Early postoperative mortality was 3.48%. There was 100% follow-up for the survivors at a mean of 4.38 ± 2.52 years. There was 1 case of thromboembolism (0.89%), 6 endocarditis (5.4%), and 9 (7.8%) conduit re-interventions. Echocardiography at discharge and last follow-up showed average peak systolic transvalvular gradients of 12.4 ± 5.1 and 18.7 ± 8.8 mm Hg, respectively. Ten patients had significant proximal anastomotic gradients of greater than 50 mm Hg and 4 moderate conduit regurgitations. Survival was 96.52%. No valve degeneration was seen in 87.82% at 5 years. The only risk factor identified for valve re-intervention was conduit implantation without infundibular hood (p = 0.01 in multivariate analysis).
Conclusions: Mid-term data show that Freestyle valves are well suited for pulmonary valve replacement in adults in both categories. The surgical technique used in valve implantation is important to ensure conduit durability. These results and accessibility to the Freestyle valve make this an acceptable alternative to homografts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.athoracsur.2015.05.006 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!