A novel implant-free tibial pull-press-fixation for ACL reconstruction.

Arch Orthop Trauma Surg

Klinikum Schaumburg, Kreiskrankenhaus Stadthagen, Am Krankenhaus 1, 31655, Stadthagen, Germany.

Published: November 2015

Introduction: Extracortical fixation techniques in anterior cruciate ligament reconstruction bear the risk of tunnel enlargement, while close-to-aperture fixations often show lower failure loads. The purpose for this study was to investigate the biomechanical benefits of a novel implant-free combination of an extra-cortical and close-to-aperture fixation.

Materials And Methods: Quadrupled human cadaveric semitendinosus tendons were fixed to 30 porcine tibiae with either a cannulated interference screw (I), an implant-free post-fixation (S), or a novel pull-press fixation (P). Specimens were cyclically loaded 20 times between 20 and 60 N followed by 500 cycles with 60-200 N, followed by a load-to-failure test with 1 mm/s.

Results: The mean elongation of the tendons in the P-group during the 500 cycles between 60 and 200 N was significantly lower (5.69 ± 2.16 mm) compared to 9.20 ± 3.21 mm in S-group and 9.37 ± 3.1 mm in the I-group (p < 0.05). The mean maximum load-to-failure was significantly higher in the P-group (728.2 ± 76.4 N) compared to 476.4 ± 68.8 N in the S-group and 625.9 ± 82.5 N in the I-group (p < 0.05). Stiffness of the constructs in the P-group was significantly higher (121.7 ± 44.9 N/mm) compared to 46.2 ± 17.7 N/mm in the S- and 72.8 ± 29.8 N/mm in the I-group (p < 0.03).

Conclusions: This study indicates superior biomechanical properties of a novel implant-free tibial pull-press fixation to conventional implant-free and close-to-aperture interference screw fixations in terms of cyclic elongation and maximum load-to-failure.

Level Of Evidence: Not applicable, basic science study.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00402-015-2293-8DOI Listing

Publication Analysis

Top Keywords

novel implant-free
12
implant-free tibial
8
interference screw
8
pull-press fixation
8
500 cycles
8
i-group 005
8
novel
4
tibial pull-press-fixation
4
pull-press-fixation acl
4
acl reconstruction
4

Similar Publications

A biomechanical analysis of four medial patellofemoral ligament reconstruction techniques.

Clin Biomech (Bristol)

November 2024

Department of Orthopedic Surgery, SUNY Upstate Medical University, 750 E. Adams St., Syracuse, New York 13210, USA. Electronic address:

Background: Implant free medial patellofemoral ligament reconstruction is an increasingly popular technique to address patellar instability. There remains sparse literature on the biomechanical differences between the most common femoral attachment fixation points including Schottle's point, superficial medial collateral ligament, and adductor magnus tendon. This study compared these reconstruction techniques as well as a novel triangle reconstruction integrating both superficial medial collateral ligament and adductor magnus tendon.

View Article and Find Full Text PDF

Is combined use of radiofrequency ablation and balloon dilation the future of interatrial communications?

Expert Rev Cardiovasc Ther

November 2022

Department of Structural Heart Disease, Fuwai Hospital, 100037, Beijing, China.

Introduction: Personalized and stable interatrial communication is an important palliative therapy for patients with heart failure. However, this remains a technically challenging task.

Areas Covered: In the past decades, substantial advancements in atrial septostomy for the creation of controllable and durable interatrial communication have been made, and numerous novel devices and techniques are in various stages of development.

View Article and Find Full Text PDF

Background: Autologous fat grafting has recently gained popularity in breast and reconstructive procedures.

Objectives: The aim of this paper was to describe a tricomposite tuberous breast reconstruction that comprises matrix dissociation through extensive tunnelization, tissue recruitment with loops, and autologous fat transfer. This approach, called "matrix modeling," was implemented by the power-assisted liposuction, loops, and lipofilling (PALLL) technique as a method to expand the lower pole, reshape the breast, and increase breast volume.

View Article and Find Full Text PDF

Dual-energy (DE) decomposition has been adopted in orthopedic imaging to measure bone composition and visualize intraarticular contrast enhancement. One of the potential applications involves monitoring of callus mineralization for longitudinal assessment of fracture healing. However, fracture repair usually involves internal fixation hardware that can generate significant artifacts in reconstructed images.

View Article and Find Full Text PDF

Background: Simple tenotomy and anchor tenodesis are commonly used in treatment of long biceps tendon (LHB) pathologies. The tenotomy can result in biceps distalization or cosmetic deformities. A novel loop Tenodesis Technique (LTD) could prevent a distalization of the biceps muscle without the risk of implant associated complications.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!