Restoration of mechanical axis in total knee arthroplasty (TKA) is correlated with improved implant survivorship. We assessed the accuracy and required surgical time using a hand-held accelerometer-based navigation system for TKA. Data collected on 53 patients included assembly, resection, and tourniquet times. Implant alignment and mechanical axis were measured on radiographs. Femoral alignment was 0.29° ± 2.2° varus. Tibial alignment was 0.09° ± 1.4° valgus. Postoperative mechanical axis was 0.2° ± 2.1° varus. Malalignment rates for the femur, tibia, and axis were 13%, 3.8%, and 17%, respectively. Average time for pinning and navigating was 3.6 minutes for the femur and 2.6 minutes for the tibia; mean tourniquet time was 62 minutes. This navigation system accurately re-established mechanical axis without increasing surgical time.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arth.2015.05.055 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!