Background And Study Aims: A new 25-gauge (G) endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) device (EchoTip ProCore; Cook Medical, Bloomington, Indiana, USA) has been developed, which features a hollowed-out reverse bevel to trap core samples. However, data on the differences between the diagnostic yield of the 25G EchoTip ProCore and that of a 22G standard needle are limited.

Patients And Methods: This pilot study included 27 patients referred during an 11-month period for EUS-FNA of pancreatic masses and enlarged lymph nodes adjacent to the upper gastrointestinal tract. Each lesion was punctured once by both a 25G EchoTip ProCore needle and a 22G standard needle (EchoTip; Cook Medical) with capillary sampling. Blinded histocytologic analyses were conducted. The final diagnosis was based on FNA findings of malignant cells, pathologic analysis of the surgical specimen, and/or radiologic and clinical follow-up of at least 7 months.

Results: A total of 28 EUS-FNA procedures targeting masses of the pancreas (n = 19) and lymph nodes (n = 9) were performed. No complications were encountered. Single-pass sensitivity rates for pancreatic and lymph node malignancy were equal for the needle types: 89.5 % (95 %CI 66.82 - 98.39) and 66 % (95 %CI 24.1 - 94), respectively. There were no significant differences between the needles in terms of EUS visualization (P = 0.125), amount of blood contamination (P = 0.705), macroscopic quantity of the material (P = 0.858), quality of the cytology (P = 0.438), and adequacy and accuracy of the cell block material (P = 0.220).

Conclusions: Both needles were safe and successful in terms of a high diagnostic yield, with similar histocytologic results. The results of this study were presented at Digestive Disease Week (DDW) 2014, Chicago, Illinois. This trial was registered at ClinicalTrials.gov (B027201316271).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4424873PMC
http://dx.doi.org/10.1055/s-0034-1390889DOI Listing

Publication Analysis

Top Keywords

echotip procore
12
cook medical
8
diagnostic yield
8
25g echotip
8
22g standard
8
standard needle
8
lymph nodes
8
needle
6
25-gauge histology
4
histology needle
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!