Ultrasound shear wave velocity in skeletal muscle: A reproducibility study.

Diagn Interv Imaging

Department of radiology, hôpital Édouard-Herriot, hospices civils de Lyon, 5, place d'Arsonval, 69437 Lyon cedex 03, France; Inserm UMR 1033, Pavillion F, hôpital Édouard-Herriot, 5, place d'Arsonval, 69437 Lyon cedex 03, France. Electronic address:

Published: January 2016

Purpose: The purpose of the study was threefold: to assess the reliability of shear wave velocities (SWV) measurements in normal skeletal muscles; to evaluate intra- and inter-operator reproducibility of measurements for a specific site of the muscle and for the mean value in the whole muscle.

Materials And Methods: Two sets of measurements were performed at three weeks intervals of each other on 16 volunteers by two radiologists on medial gastrocnemius and tibialis anterior muscles. Each muscle was evaluated in 5 different sites, with three measurements for each site in the transverse and longitudinal planes. Reliability of SWV measurements was assessed by means of intraclass correlation coefficient (ICC).

Results: Reliability of the three independent SWV measurements was excellent, slightly better in the longitudinal plane. Inter/intra-operator reproducibility per site was fair to good in the longitudinal plane and poor to fair in the transverse plane. For global values of the whole muscle, ICC showed good agreement in the longitudinal plane and fair agreement in the transverse plane.

Conclusion: Quantitative SWV measurements are reliable when performed in rigorous conditions. In conditions that mirror clinical practice, inter/intra-operator reproducibility is moderate, better for longitudinal compared to transverse plane.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diii.2015.05.010DOI Listing

Publication Analysis

Top Keywords

swv measurements
16
longitudinal plane
12
shear wave
8
better longitudinal
8
inter/intra-operator reproducibility
8
transverse plane
8
measurements
7
longitudinal
5
plane
5
ultrasound shear
4

Similar Publications

[Value of comprehensive evaluation of bladder structure and function in children by multi-modal ultrasound].

Zhonghua Yi Xue Za Zhi

January 2025

Department of Urology, Henan Joint International Pediatric Urodynamic Laboratory, the First Affiliated Hospital of Zhengzhou University, Zhengzhou450002, China.

To comprehensively evaluate the structure and function of the bladder wall in children using multi-modal ultrasound (MMU). A total of 110 children without urinary symptoms admitted to the First Affiliated Hospital of Zhengzhou University from January 2022 to June 2023 were included prospectively. MMU was performed and the vesical volume (VV), bladder wall thickness (BWT), shear wave velocity (SWV) of the anterior wall, resistance index (RI) and vascularization index (VI) were measured in the filled and emptied bladder respectively, and the ultrasound bladder compliance (UBC) was also calculated.

View Article and Find Full Text PDF

Objectives: To evaluate the value of ultrasound (US) and shear wave velocity (SWV) to assess muscle in postmenopausal women with osteosarcopenia (OSP).

Methods: This study included 145 postmenopausal women, comprising 115 osteopenia/osteoporosis participants without sarcopenia (OP alone) and 30 OSP participants. All received the evaluation of bone mineral density (BMD), appendicular skeletal muscle mass index (ASMI), handgrip strength, calf circumference, 6-meter walking speed, and 5-time chair stand test.

View Article and Find Full Text PDF

Background: When performing the Upper Limb Neurodynamic Test 1 (ULNT1), the order of joint movement can be varied to place more stress onto certain nerve segments. However, the mechanisms underlying this phenomenon are still unclear. This study aimed to analyze the differences in the stiffness of the median nerve (MN) and the brachial plexus (BP) using ultrasound shear wave elastography during three sequences of the ULNT1: standard (ULNT1-STD), distal-to-proximal (ULNT1-DIST), and proximal-to-distal (ULNT1-PROX).

View Article and Find Full Text PDF

Mesoporous silica nanoparticles as sensitizers: A novel approach to enhancing shear wave elastography in liver stiffness measurement.

Biomater Adv

January 2025

Department of Ultrasound, Xijing Hospital, Fourth Military Medical University, No.127 Changle West Rd, 710032 Xi'an, Shaanxi, China. Electronic address:

Purpose: The objective of this study is to elucidate the sensitizing effect of mesoporous silica nanoparticles (MSNs) on shear wave elastography (SWE) and to investigate the potential application of MSNs as a sensitizer to enhance the sensitivity of SWE in the diagnosis of metabolic-associated steatohepatitis (MASH).

Materials And Methods: The in vitro gelatin models with varying ratios were assessed using SWE to identify the gelatin ratio that most closely approximates with human liver stiffness. Following the characterization of the dispersion properties of MSNs, in vitro models incorporating MSNs of different particle sizes were developed.

View Article and Find Full Text PDF

Defining Normal Enthesis Stiffness Range in A Healthy Adult Population.

Br J Radiol

December 2024

Rheumatology Unit, Department of Internal Medicine, Faculty of Medicine, Universiti Teknologi MARA, 47000, Sungai Buloh, Malaysia.

Objectives: This study explores the correlation between volunteer demographics with enthesis stiffness and intra and inter -observer agreements using shear wave elastography (SWE).

Methods: 98 healthy volunteers were recruited. SWE was performed on quadriceps, suprapatellar, infrapatellar, and Achilles entheses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!