Currently available near infrared spectroscopy (NIRS) devices are unable to discriminate between arterial and venous blood, a potential source of artifact. The purpose of this study was to test the hypothesis that oscillations in NIR signals at the respiratory and cardiac frequency could be attributed to venous and arterial blood, respectively, and thereby isolated. After written informed consent was obtained, a two-wavelength NIRS device was placed over the left frontal cortex in 20 volunteers. After 5 min of unimpeded spontaneous ventilation, an impedance threshold device (ITD, average resistance-7 cm H2O) was applied and an additional two minutes of data recorded. Tissue saturation (StO2) calculated at the ventilatory and cardiac frequencies was compared to non-pulsatile StO2, before and after application of the ITD using spectral peak and power algorithms. The ITD increased non-pulsatile cerebral saturation by 3.6 %. The ITD had no discernable effect on pulsatile estimates of StO2 at either the ventilatory or cardiac frequencies. StO2 estimated at the NIRS spectral peak from 0.75 to 1.75 Hz was 24 % higher than non-pulsatile StO2 (p = 0.0013). There were no other significant differences between pulsatile and non-pulsatile algorithms in the estimation of StO2. In 64 % of cases, both the low (ventilator) and high (cardiac) frequency estimates of StO2 were either both larger or both smaller than non-pulsatile StO2, suggesting that they were interrogating the same vascular bed. Frequency domain analysis cannot reliably separate NIRS waveforms into arterial and venous components.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10877-015-9729-0DOI Listing

Publication Analysis

Top Keywords

non-pulsatile sto2
12
frequency domain
8
domain analysis
8
infrared spectroscopy
8
impedance threshold
8
threshold device
8
arterial venous
8
cardiac frequency
8
sto2
8
ventilatory cardiac
8

Similar Publications

Mitochondrial Oxygenation During Cardiopulmonary Bypass: A Pilot Study.

Front Med (Lausanne)

July 2022

Department of Anesthesiology, Erasmus Medical Center, Erasmus University Rotterdam, Rotterdam, Netherlands.

Objective: Adequate oxygenation is essential for the preservation of organ function during cardiac surgery and cardiopulmonary bypass (CPB). Both hypoxia and hyperoxia result in undesired outcomes, and a narrow window for optimal oxygenation exists. Current perioperative monitoring techniques are not always sufficient to monitor adequate oxygenation.

View Article and Find Full Text PDF

Currently available near infrared spectroscopy (NIRS) devices are unable to discriminate between arterial and venous blood, a potential source of artifact. The purpose of this study was to test the hypothesis that oscillations in NIR signals at the respiratory and cardiac frequency could be attributed to venous and arterial blood, respectively, and thereby isolated. After written informed consent was obtained, a two-wavelength NIRS device was placed over the left frontal cortex in 20 volunteers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!