An emergency bioassay method for (210)Po in urine.

Appl Radiat Isot

Canadian Nuclear Laboratories (CNL) Radiological Protection Research and Instrumentation Branch, Chalk River, ON, Canada K0J 1J0. Electronic address:

Published: September 2015

A rapid method was developed to efficiently measure (210)Po in urine samples in an emergency situation. Polonium-210 in small urine samples (10 mL) was spontaneously deposited on a stainless steel disc in 1 M HCl at room temperature for 4 h in a polyethylene bottle. The metallic disc was then counted for 4 h by alpha spectrometry. The developed method allowed the preparation of large sample batch in a short time. The method meets the requirements for an emergency bioassay procedure.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2015.06.013DOI Listing