Advanced oxidation protein products (AOPP) are reportedly elevated in the plasma of patients with a number of diseases, including diabetes mellitus, that involve oxidative stress. However, the accurate measurement of AOPP in human plasma is hampered by the formation of a precipitate following the addition of potassium iodide and glacial acetic acid according to the published assay procedure. Here we describe a modification of the AOPP assay which eliminates interference by precipitation and provides a robust, reliable, and reproducible protocol for the measurement of iodide oxidising capacity in plasma samples (intra-assay CV 1.7-5.3%, interassay CV 5.3-10.5%). The improved method revealed a significant association of AOPP levels with age (p < 0.05) and hypertension (p = 0.01) in EDTA-anticoagulated plasma samples from 52 patients with diabetes and 38 nondiabetic control subjects, suggesting a possible link between plasma oxidising capacity and endothelial and/or vascular dysfunction. There was no significant difference between AOPP concentrations in diabetic (74.8 ± 7.2 μM chloramine T equivalents) and nondiabetic (75.5 ± 7.0 μM chloramine T equivalents) individuals.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4465816 | PMC |
http://dx.doi.org/10.1155/2015/496271 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!