Objective: To compare the performance of different types of detectors in breast cancer detection.

Methods: A mammography image set containing subtle malignant non-calcification lesions, biopsy-proven benign lesions, simulated malignant calcification clusters and normals was acquired using amorphous-selenium (a-Se) detectors. The images were adapted to simulate four types of detectors at the same radiation dose: digital radiography (DR) detectors with a-Se and caesium iodide (CsI) convertors, and computed radiography (CR) detectors with a powder phosphor (PIP) and a needle phosphor (NIP). Seven observers marked suspicious and benign lesions. Analysis was undertaken using jackknife alternative free-response receiver operating characteristics weighted figure of merit (FoM). The cancer detection fraction (CDF) was estimated for a representative image set from screening.

Results: No significant differences in the FoMs between the DR detectors were measured. For calcification clusters and non-calcification lesions, both CR detectors' FoMs were significantly lower than for DR detectors. The calcification cluster's FoM for CR NIP was significantly better than for CR PIP. The estimated CDFs with CR PIP and CR NIP detectors were up to 15% and 22% lower, respectively, than for DR detectors.

Conclusion: Cancer detection is affected by detector type, and the use of CR in mammography should be reconsidered.

Key Points: The type of mammography detector can affect the cancer detection rates. CR detectors performed worse than DR detectors in mammography. Needle phosphor CR performed better than powder phosphor CR. Calcification clusters detection is more sensitive to detector type than other cancers.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4691226PMC
http://dx.doi.org/10.1007/s00330-015-3885-yDOI Listing

Publication Analysis

Top Keywords

cancer detection
16
calcification clusters
12
detectors
11
breast cancer
8
detection rates
8
types detectors
8
image set
8
non-calcification lesions
8
benign lesions
8
radiography detectors
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!