Excreta Sampling as an Alternative to In Vivo Measurements at the Hanford Site.

Health Phys

*1835 Terminal Drive, Ste 200, MSIN B1-60, Richland, WA 99354.

Published: August 2015

The capabilities of indirect radiobioassay by urine and fecal sample analysis were compared with the direct radiobioassay methods of whole body counting and lung counting for the most common radionuclides and inhalation exposure scenarios encountered by Hanford workers. Radionuclides addressed by in vivo measurement included 137Cs, 60Co, 154Eu, and 241Am as an indicator for plutonium mixtures. The same radionuclides were addressed using gamma energy analysis of urine samples, augmented by radiochemistry and alpha spectrometry methods for plutonium in urine and fecal samples. It was concluded that in vivo whole body counting and lung counting capability should be maintained at the Hanford Site for the foreseeable future, however, urine and fecal sample analysis could provide adequate, though degraded, monitoring capability for workers as a short-term alternative, should in vivo capability be lost due to planned or unplanned circumstances.

Download full-text PDF

Source
http://dx.doi.org/10.1097/HP.0000000000000304DOI Listing