Intraoperative neuromonitoring (IONM) has been commercially available for approximately 15 years and is highly predictive in thyroid gland surgery concerning either postoperative vocal fold mobility in the case of an intact signal for muscle action electromyogram (EMG, > 99 % right negative) or vocal fold dysfunction in the case of loss of signal (> 70 % right positive). The use of IONM improves the intraoperative identification of recurrent laryngeal nerve function and due to the high predictive value with respect to the expected vocal cord function the result of IONM has to be integrated into the surgical concept of thyroidectomy. Unilateral loss of function of the recurrent laryngeal nerve cannot be completely avoided despite correct application of IONM; however, bilateral vocal fold palsy can be safely avoided when contralateral surgery is cancelled after a loss of signal occurs during resection of the first side in planned bilateral surgery (alternative strategy). Patients have to be informed preoperatively about the limitations of IONM and potential strategy changes during planned bilateral surgery. Surgeons should apply IONM according to the published current recommendations and by selecting a risk-oriented intraoperative strategy in the case of loss of signal from the recurrent laryngeal nerve.