Objective: In this paper, we assess the feasibility of using two detectors in a slot-scanning digital radiography system to acquire images for measuring bone mineral density (BMD) of the middle phalanx of the middle finger using dual-energy X-ray absorptiometry (DXA).
Methods: Simulations were used to evaluate the spectral separation of the low- and high-energy spectra and detective quantum efficiency was used for assessing image quality. Scan parameters were chosen to optimize spectral separation, image quality, and radiation dose. We introduce the measurement of volumetric BMD (vBMD) using basis material decomposition. We assess the accuracy of our methods by comparing measurements taken using bone images against reference data derived from subsequent incineration of the bones. In vivo scans were conducted to evaluate the system precision (repeatability) and agreement with a clinical densitometer.
Results: Average errors for bone mineral content (BMC), areal BMD (aBMD), and vBMD were 4.85%, 5.49%, and 12.77%, respectively. Our system had good agreement with a clinical densitometer based on concordance correlation coefficient values of 0.92 and 0.98 for aBMD and BMC, respectively. Precision studies yielded coefficient of variation (CV) values of 1.35% for aBMD, 1.48% for BMC, and 1.80% for vBMD. The CV values of all measurements were within 2%, indicating that the methods have clinically acceptable precision.
Conclusion: We conclude that our techniques yield bone measurements with high accuracy, clinically acceptable precision, and good agreement with a clinical densitometer.
Significance: We have shown the clinical potential of phalangeal DXA measurements of aBMD and vBMD on a slot-scanning digital radiography system.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1109/TBME.2015.2447575 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!