Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: This study aimed to assess 6-month outcomes in patients with implantable cardioverter-defibrillators (ICDs) undergoing renal sympathetic denervation (RSD) for refractory ventricular arrhythmias (VAs).
Background: ICDs are generally indicated for patients at high risk of malignant VAs. Sympathetic hyperactivity plays a critical role in the development, maintenance, and aggravation of VAs.
Methods: A total of 10 patients with refractory VA underwent RSD. Underlying conditions were Chagas disease (n = 6), nonischemic dilated cardiomyopathy (n = 2), and ischemic cardiomyopathy (n = 2). Information on the number of ventricular tachycardia (VT)/ventricular fibrillation (VF) episodes and device therapies (antitachycardia pacing/shocks) in the previous 6 months as well as 1 and 6 months post-treatment was obtained from ICD interrogation.
Results: The median number of VT/VF episodes/antitachycardia pacing/shocks 6 months before RSD was 28.5 (range 1 to 106)/20.5 (range 0 to 52)/8 (range 0 to 88), respectively, and was reduced to 1 (range 0 to 17)/0 (range 0 to 7)/0 (range 0 to 3) at 1 month and 0 (range 0 to 9)/0 (range 0 to 7)/0 (range 0 to 3) at 6 months afterward, respectively. There were no major procedure-related complications. Two patients experienced sustained VT within the first week; in both cases, no further episodes occurred during follow-up. Two patients were nonresponders: 1 with persistent idioventricular rhythm and 1 with multiple renal arteries and incomplete ablation. Three patients died during follow-up. None of the deaths was attributed to VA.
Conclusions: In patients with ICDs and refractory VAs, RSD was associated with reduced arrhythmic burden with no procedure-related complications. Randomized controlled trials investigating RSD for treatment of refractory VAs in patients with increased sympathetic activity are needed.