Background: Cardiac CT is an excellent tool for evaluating the anatomy of a secundum atrial septal defect (ASD). However, a comprehensive assessment of its usefulness, including measurement of the pulmonary to systemic blood flow ratio in secundum ASD patients, has not been performed.

Objective: Therefore, this study was designed to evaluate the usefulness of CT for assessing the hemodynamics of secundum ASD in adults compared with transesophageal echocardiography (TEE), transthoracic echocardiography, and invasive catheterization.

Methods: Fifty adult patients with secundum ASD were enrolled. Cardiac CT scans (128-slice multidetector CT instrument) were acquired. These were followed by 2-dimensional reconstruction of the secundum ASDs to determine the defect size, the rim length between the outer edge of the defect, and the pulmonary to systemic blood flow (Qp/Qs) ratio.

Results: The maximum sizes of the secundum ASDs derived from CT and TEE studies were comparable (21.2 ± 8.0 vs. 20.0 ± 7.3 mm; P = .41; r = 0.960; P < .001). The rim lengths for the aortic, mitral, and tricuspid valves; the inferior vena cava; and posterior atrium were also comparable between CT and TEE measurements. The mean Qp/Qs ratio that was derived from CT measurements was comparable with that found by invasive catheterization (2.3 ± 0.7 vs. 2.3 ± 0.8; P = .73; r = 0.786; P < .001).

Conclusion: Cardiac CT is feasible for assessing pathology and the severity of secundum ASD in adults.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcct.2015.04.007DOI Listing

Publication Analysis

Top Keywords

secundum asd
16
comprehensive assessment
8
atrial septal
8
pulmonary systemic
8
systemic blood
8
blood flow
8
asd adults
8
secundum asds
8
secundum
7
asd
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!