[Imaging evaluation on adaptability of proximal humeral anatomy after shoulder replacement with individualized shoulder prosthesis].

Zhonghua Yi Xue Za Zhi

Department of Orthopedics, Sports Injuries Institute of Chongqing, Orthopedic Center of PLA, Southwest Hospital, Third Military Medical University, Chongqing 400038, China.

Published: March 2015

Objective: Modern shoulder prosthesis has evolved through four generations. And the fourth generation technology has a core three-dimensional design of restoring 3D reconstruction of proximal humeral anatomy. Thus a new shoulder prosthesis is developed on the basis of the technology of 3D prosthesis. Assessment of whether shoulder prosthesis can restore individualized reconstruction of proximal humeral anatomy is based on the adaptability of proximal humeral anatomy. To evaluate the adaptability of proximal humeral anatomy through measuring the parameters of proximal humeral anatomy after shoulder replacement with individualized shoulder prosthesis and compare with normal data.

Methods: The parameters of proximal humeral anatomy were analyzed and evaluated for a total of 12 cases undergoing shoulder replacement with individualized shoulder prosthesis. The relevant anatomical parameters included neck-shaft angle (NSA), retroversion angle (RA), humeral head height (HH) and humeral head diameter (HD). And the anatomical parameters were compared with the data from normal side.

Results: All underwent shoulder replacement with individualized shoulder prosthesis. The postoperative parameters of proximal humeral anatomy were compared with those of normal side. And the difference of NSA was < 3°, RA < 3°, HH < 3 mm and HD < 2 mm. And paired-sample t test was used to study the parameters of proximal humeral anatomy between postoperative and normal side. The normal and postoperative NSA was (140.2 ± 6.8)° and (139.5 ± 6.6)° respectively, RA (34.4 ± 3.3)° and (33.8-3.1)°, HH (15.3 ± 2.1) mm and (14.6+0.9) mm, HW (42.2 ± 2.82) mm and (41.8 ± 2.33) mm respectively. No significant difference existed between two groups (P > 0.05).

Conclusion: Individualized shoulder prosthesis has excellent adaptability to shoulder. All core parameters are freely adjustable and specification models may be optimized. With matching tools, individualized shoulder prosthesis improves the accuracy and reliability in shoulder replacement.

Download full-text PDF

Source

Publication Analysis

Top Keywords

proximal humeral
36
humeral anatomy
36
shoulder prosthesis
32
individualized shoulder
24
shoulder replacement
20
replacement individualized
16
parameters proximal
16
shoulder
15
adaptability proximal
12
anatomy shoulder
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!