Background: To evaluate and quantify the impact of a novel image-based motion correction technique in myocardial T2 mapping in terms of measurement reproducibility and spatial variability.

Methods: Twelve healthy adult subjects were imaged using breath-hold (BH), free breathing (FB), and free breathing with respiratory navigator gating (FB + NAV) myocardial T2 mapping sequences. Fifty patients referred for clinical CMR were imaged using the FB + NAV sequence. All sequences used a T2 prepared (T2prep) steady-state free precession acquisition. In-plane myocardial motion was corrected using an adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC). DICE similarity coefficient (DSC) and myocardial boundary errors (MBE) were measured to quantify the motion estimation accuracy in healthy subjects. T2 mapping reproducibility and spatial variability were evaluated in healthy subjects using 5 repetitions of the FB + NAV sequence with either 4 or 20 T2prep echo times (TE). Subjective T2 map quality was assessed in patients by an experienced reader using a 4-point scale (1-non diagnostic, 4-excellent).

Results: ARCTIC led to increased DSC in BH data (0.85 ± 0.08 vs. 0.90 ± 0.02, p = 0.007), FB data (0.78 ± 0.13 vs. 0.90 ± 0.21, p < 0.001), and FB + NAV data (0.86 ± 0.05 vs. 0.90 ± 0.02, p = 0.002), and reduced MBE in BH data (0.90 ± 0.40 vs. 0.64 ± 0.19 mm, p = 0.005), FB data (1.21 ± 0.65 vs. 0.63 ± 0.10 mm, p < 0.001), and FB + NAV data (0.81 ± 0.21 vs. 0.63 ± 0.08 mm, p < 0.001). Improved reproducibility (4TE: 5.3 ± 2.5 ms vs. 4.0 ± 1.5 ms, p = 0.016; 20TE: 3.9 ± 2.3 ms vs. 2.2 ± 0.5 ms, p = 0.002), reduced spatial variability (4TE: 12.8 ± 3.5 ms vs. 10.3 ± 2.5 ms, p < 0.001; 20TE: 9.7 ± 3.5 ms vs. 7.5 ± 1.4 ms) and improved subjective score of T2 map quality (3.43 ± 0.79 vs. 3.69 ± 0.55, p < 0.001) were obtained using ARCTIC.

Conclusions: The ARCTIC technique substantially reduces spatial mis-alignment among T2-weighted images and improves the reproducibility and spatial variability of in-vivo T2 mapping.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4465156PMC
http://dx.doi.org/10.1186/s12968-015-0141-1DOI Listing

Publication Analysis

Top Keywords

reproducibility spatial
12
myocardial mapping
12
motion correction
8
spatial variability
8
free breathing
8
fb + nav sequence
8
healthy subjects
8
myocardial
5
impact motion
4
correction reproducibility
4

Similar Publications

Purpose: This study investigates the capabilities of ultrasonography (US) in determing the stage of orbital inflammation in patients with granulomatosis with polyangiitis (GPA).

Material And Methods: The study included 24 patients (8 men and 16 women) with diffuse orbital tissue involvement in GPA. Group 1 (active stage) included nine patients, while group 2 (inactive stage) consisted of 18 patients.

View Article and Find Full Text PDF

Background: For the development and validation of dynamic treatment modalities and processes on the MR-linac, independent measurements should be performed that validate dose delivery and linac behavior at a high temporal resolution. To achieve this, a detector with both high temporal and spatial resolution is necessary.

Purpose: This study investigates the suitability of a Delta4 Phantom+ MR (Delta4) detector array for time-resolved dosimetry in the 1.

View Article and Find Full Text PDF

The Morris Water Maze (MWM) is the most commonly used assay for evaluating learning and memory in laboratory mice. Despite its widespread use, contemporary reviews have highlighted substantial methodological variation in experimental protocols and that the associated testing procedures are acutely (each trial) and chronically (testing across days) stressful; stress impairs attention, memory consolidation and the retrieval of learned information. Moreover, the interpretation of behavior within the MWM is often difficult because of wall hugging, non-spatial swim strategies, floating, and jumping off the escape platform.

View Article and Find Full Text PDF

The development and validation of a tablet-based assessment battery of general cognitive ability.

BMC Psychol

December 2024

Beijing Infinite Brain Technology Co., Ltd, Beijing, 100022, PR China.

Background: Traditional cognitive assessments, often reliant on paper-and-pencil tests and professional evaluators, suffer from subjectivity and limited result discrimination. This study introduces the Baguan Online Cognitive Assessment System (BOCAS), a tablet-based system that evaluates both general cognitive ability (GCA) and domain-specific functions across six domains: sensory-motor skills, processing speed, sustained attention, working memory, cognitive flexibility, and spatial ability.

Methods: BOCAS was validated with 151 healthy Chinese adults aged 18-40.

View Article and Find Full Text PDF

Deciphering single-cell 3D chromatin structure using scCTG.

J Chem Phys

December 2024

Beijing National Laboratory for Molecular Sciences, College of Chemistry and Molecular Engineering, Peking University, Beijing 100871, China.

Sequencing-based Hi-C technology has been widely used to study the three-dimensional structure of chromatin. More recently, the development of single-cell Hi-C technology has enabled the study of chromatin structural variations between individual cells. However, single-cell Hi-C data are often highly sparse, necessitating the use of imputation algorithms to address insufficient sampling.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!