Performance evaluation of four 25-hydroxyvitamin D assays to measure 25-hydroxyvitamin D2.

Clin Biochem

Siemens Healthcare Diagnostics Inc., 511 Benedict Ave, Tarrytown, NY 10591, USA. Electronic address:

Published: November 2015

Objectives: The ability of current immunoassays to accurately measure equimolar amounts of 25(OH)D2 and 25(OH)D3 has been recently questioned. This study determined serum 25(OH)D2, 25(OH)D3 and total serum 25(OH)D concentrations in healthy vitamin D2-supplemented subjects by isotope dilution liquid chromatography mass spectrometry (ID-LC-MS/MS); and, evaluated the ability of the Siemens, DiaSorin, Roche, and Abbott Vitamin D Total assays to monitor total serum 25(OH)D concentrations compared to an ID-LC-MS/MS method traceable to the National Institute of Standards and Technology (NIST), and that has achieved certification from the Centers for Disease Control and Prevention (CDC) Vitamin D Standardization Certification Program (VDSCP).

Design And Methods: Twenty (20) healthy adults, with no history of prior vitamin D supplementation were administered oral vitamin D2 (2400IU/day for 6months). Serum samples (140) from baseline and monthly blood draws were tested.

Results: After one month, the mean serum 25(OH)D2 concentrations rose from 0.8 to 43.6nmol/L, whereas 25(OH)D3 concentrations declined from 84.0 to 63.4nmol/L; total serum 25(OH)D concentrations rose from 86.6 to 107.0nmol/L. The overall mean bias to ID-LC-MS/MS was -7.1% for the Siemens ADVIA Centaur assay, -15.3% for the DiaSorin LIAISON assay; -8.4% for the Roche ELECSYS assay and -16.3% for the Abbott ARCHITECT assay. Correlation coefficients (r) were 0.94, 0.79, 0.74, and 0.73; the mean bias for baseline [25(OH) D3-containing] versus six-month [25(OH)D2- and 25(OH)D3-containing] samples was -13.4% and -5.7%; -3.5% and 20.3%, 9.6% and -12.1%, and 0.2% and -17.8%, respectively.

Conclusions: The bias results obtained for the Siemens ADVIA Centaur assay and Roche ELECSYS assay were slightly lower than those for the DiaSorin LIAISON assay and the Abbott ARCHITECT assay, but all 25(OH)D assays demonstrated acceptable performance.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2015.05.021DOI Listing

Publication Analysis

Top Keywords

total serum
12
serum 25ohd
12
25ohd concentrations
12
25ohd2 25ohd3
8
serum 25ohd2
8
concentrations rose
8
siemens advia
8
advia centaur
8
assay
8
centaur assay
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!