Load to failure of different titanium abutments for an internal hexagon implant.

J Prosthet Dent

Professor, Division of Restorative and Prosthetic Dentistry, The Ohio State University, College of Dentistry, Columbus, Ohio.

Published: October 2015

Statement Of Problem: Several aftermarket abutments are available for a commonly used internal hexagonal connection implant. However, their load to failure performance is unknown when compared with the manufacturer's abutment.

Purpose: The purpose of this in vitro study was to conduct a load to failure comparison of 5 different titanium abutments (manufacturer's and aftermarket) for cement-retained restorations used on an implant with an internal hexagon connection.

Material And Methods: Five implants (Tapered Screw-Vent, 4.1×11.5 mm; Zimmer Dental) were individually secured in a loading apparatus, and 3 abutment specimens of each of the 5 different titanium abutments (Atlantis, AstraTech TiDesign, Legacy Straight Contoured, Inclusive Custom, and Zimmer PSA) (n=15 total) were loaded at a 30-degree angle until fracture of the implant abutment complex. Data for load to fracture were compared with analysis of variance and a Tukey-Kramer post hoc test (α=.05).

Results: Significant differences were noted between the fracture loads of some abutment pairs; Atlantis-AstraTech TiDesign, Atlantis-Legacy Straight Contoured, AstraTech TiDesign-Legacy Straight Contoured, Inclusive Custom-AstraTech TiDesign, and Inclusive Custom-Legacy Straight Contoured (P<.05). The highest overall resistance to fracture was achieved by the Legacy Straight Contoured Abutment, which was significantly greater than all other aftermarket abutments (P<.05). Tested abutments fractured at an average of 649.17 N. The Zimmer PSA abutment was the only abutment that showed no fracture of any of the components before implant failure.

Conclusion: When comparing manufacturer's versus aftermarket brands, the manufacturer's abutment (Zimmer PSA) was the only abutment without fracture of any of the components. Aftermarket brands experienced screw fractures, which could result in further clinical prosthetic complications. The clinical implications of these findings need further investigation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.prosdent.2015.04.012DOI Listing

Publication Analysis

Top Keywords

straight contoured
16
load failure
12
titanium abutments
12
internal hexagon
8
contoured inclusive
8
load
4
failure titanium
4
abutments
4
abutments internal
4
implant
4

Similar Publications

Straightness is the basic measurement parameter in machining, and the traditional straightness measurement methods such as light gap method, table method, et al., have extremely low measurement efficiency and cannot achieve online real-time high-precision detection. Our research group has proposed a machine vision online detection based on 10 industrial camera arrays, which can obtain the surface profile straight line of the sucker rod by collecting the edge profile image of the sucker rod and performing morphological transformation.

View Article and Find Full Text PDF

The precise measurement of inner dimensions and contour accuracy is required for deep-hole parts, particularly during the manufacturing process, to monitor quality and obtain real-time error parameters. However, on-machine measurement is challenging due to the limited inner space of deep holes. This study proposes an automatic on-machine measuring apparatus for assessing inner diameter, straightness, and roundness errors.

View Article and Find Full Text PDF

Peripheral straightness leads to shape diversification during formations of entire leaves.

J Theor Biol

January 2025

International Research Organization for Advanced Science and Technology (IROAST), Kumamoto University, Kumamoto, Japan; Meiji Institute for Advanced Study of Mathematical Science (MIMS), Meiji University, Tokyo, Japan. Electronic address:

The ways to read out positional information are essential to determine final shapes in developmental processes. Relative shaping to different sizes of positional information enables robust morphogenesis; however, the same difference sometimes causes diversity. Different responses to a positional information will enable such switching of identical/diverse shapes, though detail mechanisms remain unknown.

View Article and Find Full Text PDF

Multi-Feature-Filtering-Based Road Curb Extraction from Unordered Point Clouds.

Sensors (Basel)

October 2024

Shanghai Tongke Transportation Technology Co., Ltd., Shanghai 200092, China.

Road curb extraction is a critical component of road environment perception, being essential for calculating road geometry parameters and ensuring the safe navigation of autonomous vehicles. The existing research primarily focuses on extracting curbs from ordered point clouds, which are constrained by their structure of point cloud organization, making it difficult to apply them to unordered point cloud data and making them susceptible to interference from obstacles. To overcome these limitations, a multi-feature-filtering-based method for curb extraction from unordered point clouds is proposed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!