Article Abstract

Background: Implantable cardioverter-defibrillators (ICDs) are generally considered a contraindication to magnetic resonance imaging (MRI).

Objective: The purpose of the ProMRI Phase C study, a multicenter, prospective, single-arm, nonrandomized study, was to evaluate the clinical safety of the Biotronik ProMRI Iforia ICD system during MRI.

Methods: Patients were enrolled after ICD implantation, with either a dual-chamber DR-T or single-lead VR-T DX system. Study-defined, nondiagnostic cardiac or thoracic spine MRI was performed at least 1 week after enrollment. ICDs were placed into MRI mode with ventricular fibrillation (VF) detection/therapy programmed "off" before scan and restored to non-MRI mode after scan. Interrogation was performed before, immediately after, and 1 month post-MRI. The primary end-points were (1) ventricular pacing threshold increase >0.5 V from pre-MRI to 1 month post-MRI; (2) R-wave amplitude decrease >50% from pre-MRI to 1 month post-MRI or R-wave amplitude <5 mV at 1 month post-MRI; and (3) MRI and ICD system-related serious adverse device effects.

Results: One hundred seventy patients were enrolled at 39 US centers. One hundred fifty-three patients underwent MRI (25.7% cardiac, 74.3% thoracic spine) and completed follow-up. Freedom from the primary end-points was met in all but 1 subject, in whom reduced R-wave amplitude was detected 1 month post-MRI. No serious adverse device effects occurred during the course of the study.

Conclusion: These results demonstrate the clinical safety and efficacy of the ProMRI ICD system in patients subjected to thoracic spine and cardiac MRI imaging in 1.5-T scanners.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2015.06.002DOI Listing

Publication Analysis

Top Keywords

month post-mri
12
clinical safety
8
thoracic spine
8
magnetic resonance
8
resonance imaging
8
pre-mri month
8
post-mri r-wave
8
r-wave amplitude
8
safety iforia
4
iforia implantable
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!