The authors report an unusual case of intrathecal baclofen withdrawal due to the perforation and subsequent leakage of a baclofen pump catheter in a patient with spastic cerebral palsy. A 15-year-old boy underwent an uncomplicated placement of an intrathecal baclofen pump for the treatment of spasticity due to cerebral palsy. After excellent control of symptoms for 3 years, the patient presented to the emergency department with increasing tremors following a refill of his baclofen pump. Initial evaluation consisted of radiographs of the pump and catheter, which appeared normal, and a successful aspiration of CSF from the pump's side port. A CT dye study revealed a portion of the catheter directly overlying the refill port and extravasation of radiopaque dye into the subfascial pocket anterior to the pump. During subsequent revision surgery, a small puncture hole in the catheter was seen to be leaking the drug. The likely cause of the puncture was an inadvertent perforation of the catheter by a needle during the refilling of the pump. This case report highlights a unique complication in a patient with an intrathecal baclofen pump. Physicians caring for these patients should be aware of this rare yet potential complication in patients presenting with baclofen withdrawal symptoms.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2015.2.PEDS14501DOI Listing

Publication Analysis

Top Keywords

baclofen pump
20
pump catheter
12
intrathecal baclofen
12
baclofen withdrawal
8
cerebral palsy
8
baclofen
7
catheter
7
pump
7
catheter leakage
4
leakage migration
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!