Evaluation of a new automated assay for the measurement of circulating 1,25-dihydroxyvitamin D levels in daily practice.

Clin Biochem

Unit of Hormonal Biology, Department of Biochemistry, Saint-Louis University Hospital (AP-HP), 1 avenue Claude Vellefaux, 75475 Paris cedex 10, France. Electronic address:

Published: November 2015

Background And Objective: 1,25-dihydroxyvitamin D [1,25(OH)2D] is the most important vitamin D metabolite in regulating calcium and phosphorus metabolism and bone resorption. It is measured in plasma for diagnosis and management of patients with disorders influencing vitamin D metabolism. The most common methods for 1,25(OH)2D quantification are competitive isotopic I(125) immunoassays. The purpose of this work is to compare the IDS isotopic immunoassay (IDS RIA) and the nonisotopic iSYS immunoassay (IDS iSYS) for 1,25(OH)2D measurement in human serum and in control samples obtained from the Vitamin D External Quality Assessment Scheme (DEQAS).

Methods: 1,25(OH)2D concentrations in 180 serum samples (87 females and 93 males) and in 25 samples from the DEQAS were assayed using IDS RIA and IDS iSYS methods. Both assays were performed after delipidation and immunoextraction steps. Measurement range was 9.0-348.0pmol/L and 15.6-504.0pmol/L for IDS RIA and IDS iSYS assay, respectively. 'Normal adult' reference ranges of 1,25(OH)2D were 43-168pmol/L for IDS RIA and 63-228pmol/L for IDS iSYS.

Results: The equation of the Deming regression analysis demonstrated that IDS iSYS tended to give lower 1,25(OH)2D concentrations by a mean of 20% over the tested concentration range when compared to IDS RIA. Both assays would easily meet the DEQAS goal of 80% of survey samples within 30% of the All-Laboratory Trimmed Mean "ALTM". Results of 1,25(OH)2D obtained using IDS iSYS in samples distributed by DEQAS were closely related to those by LC-MS/MS method.

Conclusions: The concentrations of 1,25(OH)2D measured by the IDS iSYS tended to be lower than those of IDS RIA in patients, but quite comparable to those of LC-MS/MS and ALTM in DEQAS samples.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2015.05.018DOI Listing

Publication Analysis

Top Keywords

ids ria
24
ids isys
24
ids
14
immunoassay ids
8
125oh2d concentrations
8
ria ids
8
isys tended
8
tended lower
8
125oh2d
7
isys
7

Similar Publications

Harmonization of commercial assays for PINP; the way forward.

Osteoporos Int

March 2020

Department of Clinical Chemistry, University of Liège, CHU Sart-Tilman, Domaine du Sart-Tilman, B-4000, Liège, Belgium.

Unlabelled: International Federation of Clinical Chemistry and Laboratory Medicine and The International Osteoporosis Foundation Joint Committee on Bone Metabolism believes that the harmonization of PINP assays is an achievable and practical goal.

Introduction: In order to examine the agreement between current commercial assays, a multi-center study was performed for PINP in serum and plasma.

Methods: The automated methods for PINP (Roche Cobas and IDS iSYS) gave similar results.

View Article and Find Full Text PDF

Comparison of two 25-hydroxyvitamin D immunoassays to liquid chromatography-tandem mass spectrometry in assessing samples from the Chinese population.

Clin Chim Acta

August 2015

National Institute of Nutrition and Health, Chinese Center for Disease Control and Prevention, Nanwei Road No. 29, Xicheng District, Beijing postcode 100050, China. Electronic address:

Background: We compared the utility of two 25-OH-D immunoassays to a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for assessing human serum samples from the Chinese population.

Methods: In total, 451 Chinese serum samples were tested using the IDS 25-Hydroxy Vitamin D EIA kits and the LC-MS/MS 25-OH-D method. Of these, 239 samples were also analysed using DiaSorin 25-Hydroxyvitamin D RIA kits.

View Article and Find Full Text PDF

Evaluation of a new automated assay for the measurement of circulating 1,25-dihydroxyvitamin D levels in daily practice.

Clin Biochem

November 2015

Unit of Hormonal Biology, Department of Biochemistry, Saint-Louis University Hospital (AP-HP), 1 avenue Claude Vellefaux, 75475 Paris cedex 10, France. Electronic address:

Background And Objective: 1,25-dihydroxyvitamin D [1,25(OH)2D] is the most important vitamin D metabolite in regulating calcium and phosphorus metabolism and bone resorption. It is measured in plasma for diagnosis and management of patients with disorders influencing vitamin D metabolism. The most common methods for 1,25(OH)2D quantification are competitive isotopic I(125) immunoassays.

View Article and Find Full Text PDF

Background: Two new immunoassay methods for aldosterone assay using automated platforms recently became available into market. The main aim of the present study is to evaluate the analytical performance of these automated direct immunoassay methods, and also to compare their analytical characteristics to those of the most popular RIA and EIA methods used in an Italian External Quality Assessment (EQA) study.

Methods: In this study analytical performances of two aldosterone immunoassays using the IDS iSYS and DiaSorin LIAISON fully automated platforms, were evaluated.

View Article and Find Full Text PDF

Background: Over the last decade, clinical interest to evaluate human 25-hydroxy-vitamin D (25[OH]D) serum levels has increased exponentially. In the present study, four chemiluminescence immunoassays (CLIA), one radioimmunoassy (RIA), and one high performance liquid chromatography (HPLC) method were compared and also with the liquid chromatography-tandem mass spectrometry (LC-MS/MS) method in view of 25(OH)D serum level determination. Methods: For the method comparison, blood samples from 133 consecutive patients were prospectively collected.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!