Objectives: The goal of this study was to investigate the durability and longevity of gelatin formulas for the production of staged ultrasound phantoms for education.

Methods: Gelatin phantoms were prepared from Knox gelatin (Kraft Foods, Northfield, IL) and a standard 10%-by-mass ordinance gelatin solution. Phantoms were durability tested by compressing to a 2-cm depth until cracking was visible. Additionally, 16 containers with varying combinations of phenol, container type, and storage location were tested for longevity against desiccation and molding. Once formulation was determined, 4 stages of phantoms from novice to clinically relevant were poured, and clinicians with ultrasound training ranked them on a 7-point Likert scale based on task difficulty, phantom suitability, and fidelity.

Results: On durability testing, the ballistic gelatin outperformed the Knox gelatin by more than 200 compressions. On longevity testing, gelatin with a 0.5% phenol concentration stored with a lid and refrigeration lasted longest, whereas containers without a lid had desiccation within 1 month, and those without phenol became moldy within 6 weeks. Ballistic gelatin was more expensive when buying in small quantities but was 7.4% less expensive when buying in bulk. The staged phantoms were deemed suitable for training, but clinicians did not consistently rank the phantoms in the intended order of 1 to 4 (44%).

Conclusions: Refrigerated and sealed ballistic gelatin with phenol was a cost-effective method for creating in-house staged ultrasound phantoms suitable for large-scale ultrasound educational training needs. Clinician ranking of phantoms may be influenced by current training methods that favor biological tissue scanning as easier.

Download full-text PDF

Source
http://dx.doi.org/10.7863/ultra.34.6.1011DOI Listing

Publication Analysis

Top Keywords

ballistic gelatin
16
gelatin
10
staged ultrasound
8
phantoms
8
ultrasound phantoms
8
knox gelatin
8
expensive buying
8
ultrasound
5
shooting sound
4
sound optimizing
4

Similar Publications

Bullet characterization using Photon-Counting detector CT: A phantom study with intact bullets.

Eur J Radiol

December 2024

Department of Radiology and Nuclear Medicine, Maastricht University Medical Centre+, PO Box 5800, 6202 AZ Maastricht, the Netherlands; Mental Health and Sciences (MHeNs) Research Institute, Maastricht University, PO Box 616, 6200 MD Maastricht, the Netherlands.

Objectives: Photon-counting detector CT (PCD-CT) is expected to substantially improve and expand CT-imaging applicability due to its intrinsic spectral capabilities, increased spatial resolution, reduced electronic noise, and improved image contrast. The current study aim is to evaluate PCD-CT efficacy in characterizing bullets based on their dimensions, shape, and material composition.

Materials And Methods: This is an observational phantom study examining 11 unfired, intact bullets of various common calibers, placed in ballistic gelatin.

View Article and Find Full Text PDF

The development of artificial skin that accurately mimics the mechanical properties of human skin is crucial for a wide range of applications, including surgical training for burn injuries, biomechanical testing, and research in sports injuries and ballistics. While traditional materials like gelatin, polydimethylsiloxane (PDMS), and animal skins (such as porcine and bovine skins) have been used for these purposes, they have inherent limitations in replicating the intricate properties of human skin. In this work, we conducted uniaxial tensile tests on freshly obtained cadaveric skin to analyze its mechanical properties under various loading conditions.

View Article and Find Full Text PDF

To reveal the penetration mechanism and present the penetration characteristics of high-speed micro-jet with injection volume larger than 0.3 mL into soft tissue, the present study conducted experimental research on high-speed water-jet penetration into ballistic gelatin. The free jet dynamics of an air-powered needle-free injector that can emit up to 1.

View Article and Find Full Text PDF

Introduction: In accordance with the Polish law, pneumatic weapons are weapons that use compressed gas to set a bullet in motion and generate a discharge energy of more than 17J. In recent years, Poland has seen an increase in crimes committed with pneumatic weapons. The research aim of the present study was to measure the velocity of selected 5.

View Article and Find Full Text PDF

Expanding bullets and ballistic gelatin - A restricted expansion experiment.

Leg Med (Tokyo)

November 2024

Research Unit of Health Sciences and Technology, Medical Research Center Oulu, Oulu University Hospital and University of Oulu, P.O. Box 8000, FI-90014 Oulu, Finland; Department of Forensic Medicine, University of Helsinki, P.O. Box 21, FI-00014 Helsinki, Finland; Forensic Medicine Unit, Finnish Institute for Health and Welfare, P.O. Box 30, FI-00271 Helsinki, Finland.

Ballistic gelatin has gained a status as standard method for terminal ballistic testing and experimenting. Variation considering the recipe and manufacturing of the blocks exists. The golden standard has been a cuboid gelatin block, dimensions varying according to the type and kinetic energy of the ammunition.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!