Background And Purpose: In traditional radiostereometric analysis (RSA), 1 segment defines both the acetabular shell and the polyethylene liner. However, inserting beads into the polyethylene liner permits employment of the shell and liner as 2 separate segments, enabling distinct analysis of the precision of 3 measurement methods in determining femoral head penetration and shell migration.
Patients And Methods: The UmRSA program was used to analyze the double examinations of 51 hips to determine if there was a difference in using the shell-only segment, the liner-only segment, or the shell + liner segment to measure wear and acetabular cup stability. The standard deviation multiplied by the critical value (from a t distribution) established the precision of each method.
Results: Due to the imprecision of the automated edge detection, the shell-only method was least desirable. The shell + liner and liner-only methods had a precision of 0.115 mm and 0.086 mm, respectively, when measuring head penetration. For shell migration, the shell + liner had a precision of 0.108 mm, which was better than the precision of the shell-only method. In both the penetration and migration analyses, the shell + liner condition number was statistically significantly lower and the bead count was significantly higher than for the other methods.
Interpretation: Insertion of beads in the polyethylene improves the precision of femoral head penetration and shell migration measurements. A greater dispersion and number of beads when combining the liner with the shell generated more reliable results in both analyses, by engaging a larger portion of the radiograph.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4564778 | PMC |
http://dx.doi.org/10.3109/17453674.2015.1054255 | DOI Listing |
Indian J Orthop
December 2024
Trauma & Orthopaedics, Hinchingbrooke Hospital, North-West Anglia NHS Foundation Trust, Huntingdon, UK.
Total hip arthroplasty stands as a milestone in surgical success for alleviating hip arthritis-related pain and enhancing patients' quality of life. While complications persist, advancements like dual-mobility articulation aims to mitigate risks. This manuscript presents an unprecedented unique complication in a 60-year-old woman with a revision dual mobility THA.
View Article and Find Full Text PDFPLoS Med
November 2024
University of Bristol Medical School, Translational Health Sciences, Musculoskeletal Research Unit, Bristol, United Kingdom.
Background: The risk of re-operation, otherwise known as revision, following primary hip replacement depends in part on the prosthesis implant materials used. Current performance evidences are based on a broad categorisation grouping together different materials with potentially varying revision risks. We investigated the revision rate of primary total hip replacement (THR) reported in the National Joint Registry by specific types of bearing surfaces used.
View Article and Find Full Text PDFArthroplast Today
December 2024
Department of Orthopaedic Surgery, University of Washington, Seattle, WA.
Dual-mobility bearings are being increasingly utilized in total hip arthroplasty. Contemporary modular designs often feature inner cobalt-chromium liners that are seated in outer titanium acetabular shells. However, mating of these 2 dissimilar metals may lead to complications.
View Article and Find Full Text PDF3D Print Med
September 2024
The Royal National Orthopaedic Hospital, Stanmore, HA7 4LP, UK.
Background: The Trident II Tritanium acetabular shell is additively manufactured (3D printed), based on the established Trident 'I' Tritanium shell, produced using conventional methods; this study characterised their differences.
Methods: We obtained 5 Trident I (T1) and 5 Trident II (T2) shells sized 52 mm, 54 mm (n = 3) and 60 mm. We measured their: mass, shell-liner engaging surface roughness, roundness, wall thickness, the depth of the bone-facing porous layer, porosity, and the number, volume and location of structural voids.
ChemistryOpen
December 2024
Glass Research Division, Institute of Glass & Ceramic Research and Testing, Bangladesh Council of Scientific and Industrial Research (BCSIR), Dhaka, 1205, Bangladesh.
The study aims to synthesize nano-crystallite TSP using renewable, low-cost, waste marine mollusk from three different species such as Babylonia japonica, Oliva sayana, and Conasprella bermudensis. The molar ratio of phosphate to calcium in triple superphosphate [TSP, Ca(HPO).HO] significantly impacts its properties and fertilizer performance, in this case, we kept the ratio to 2.