The goal of this study was to develop and improve an infrared (IR) navigation system to deliver light dose uniformly during intracavitory PDT by tracking the movement of the light source and providing real-time feedback on the light fluence rate on the entire cavity surface area. In the current intrapleural PDT protocol, several detectors placed in selected locations in the pleural cavity monitor the light doses. To improve the delivery of light dose uniformity, an IR camera system is used to track the motion of the light source as well as the surface contour of the pleural cavity. Monte-Carlo simulation is used to improve the calculation algorithm for the effect of light that undergoes multiple scattering along the surface in addition to an improvement of the direct light calculation using an improved model that accounts for the anisotropy of the light from the light source.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4437627 | PMC |
http://dx.doi.org/10.1117/12.2080110 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!