The precision of textural analysis in (18)F-FDG-PET scans of oesophageal cancer.

Eur Radiol

Division of Imaging Sciences and Biomedical Engineering, King's College London, London, UK.

Published: September 2015

Objectives: Measuring tumour heterogeneity by textural analysis in (18)F-fluorodeoxyglucose positron emission tomography ((18)F-FDG PET) provides predictive and prognostic information but technical aspects of image processing can influence parameter measurements. We therefore tested effects of image smoothing, segmentation and quantisation on the precision of heterogeneity measurements.

Methods: Sixty-four (18)F-FDG PET/CT images of oesophageal cancer were processed using different Gaussian smoothing levels (2.0, 2.5, 3.0, 3.5, 4.0 mm), maximum standardised uptake value (SUVmax) segmentation thresholds (45%, 50%, 55%, 60%) and quantisation (8, 16, 32, 64, 128 bin widths). Heterogeneity parameters included grey-level co-occurrence matrix (GLCM), grey-level run length matrix (GLRL), neighbourhood grey-tone difference matrix (NGTDM), grey-level size zone matrix (GLSZM) and fractal analysis methods. The concordance correlation coefficient (CCC) for the three processing variables was calculated for each heterogeneity parameter.

Results: Most parameters showed poor agreement between different bin widths (CCC median 0.08, range 0.004-0.99). Segmentation and smoothing showed smaller effects on precision (segmentation: CCC median 0.82, range 0.33-0.97; smoothing: CCC median 0.99, range 0.58-0.99).

Conclusions: Smoothing and segmentation have only a small effect on the precision of heterogeneity measurements in (18)F-FDG PET data. However, quantisation often has larger effects, highlighting a need for further evaluation and standardisation of parameters for multicentre studies.

Key Points: • Heterogeneity measurement precision in (18) F-FDG PET is influenced by image processing methods. • Quantisation shows large effects on precision of heterogeneity parameters in (18) F-FDG PET/CT. • Smoothing and segmentation show comparatively smaller effects on precision of heterogeneity parameters.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-015-3681-8DOI Listing

Publication Analysis

Top Keywords

precision heterogeneity
16
smoothing segmentation
12
heterogeneity parameters
12
ccc median
12
effects precision
12
textural analysis
8
oesophageal cancer
8
heterogeneity
8
18f-fdg pet
8
image processing
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!