Dry phantom for magnetoencephalography -Configuration, calibration, and contribution.

J Neurosci Methods

Applied Electronics Laboratory, Kanazawa Institute of Technology, 3 Amaike, Kanazawa, Ishikawa 920-1331, Japan.

Published: August 2015

Background: An artificial object that imitates human brain activity is called "phantom" and is used for evaluation of magnetoencephalography (MEG) systems. The accuracy of the phantom itself had not been guaranteed in the previous studies, although role of the phantom is to evaluate the accuracy of MEG measurement. The purposes of this paper are to develop a novel MEG phantom that can be calibrated and to demonstrate the advantages of the calibrated phantoms.

New Method: We proposed and fabricated a practical dry phantom that is composed of 50 isosceles-triangle coils based on Ilmoniemi's model. This phantom was calibrated based on three-dimensional measurement of the current paths in the phantom and on numerical calculations.

Results: The calibrated positions of the equivalent current dipoles (ECDs) shifted 0.83mm, on average, from the designed positions. The uncertainties of the calibrated ECDs were also evaluated, by combining the uncertainties which could reasonably be attributed to them.

Comparison With Existing Method(s): Furthermore, we demonstrated performance of the developed phantom through experimental evaluation of an MEG system. The results of this evaluation differed from those obtained using an uncalibrated phantom. Moreover, the calibrated phantom can provide detailed information regarding the uncertainty of the measurement and also the uncertainty of the phantom itself.

Conclusions: A more appropriate evaluation of MEG measurements can be achieved using a calibrated phantom.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jneumeth.2015.05.004DOI Listing

Publication Analysis

Top Keywords

phantom calibrated
12
phantom
11
dry phantom
8
evaluation meg
8
calibrated phantom
8
calibrated
7
meg
5
phantom magnetoencephalography
4
magnetoencephalography -configuration
4
-configuration calibration
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!