A technical support tool for joint range of motion determination in functional diagnostics - an inter-rater study.

J Occup Med Toxicol

Institute of Occupational and Social Medicine, Medical Faculty, RWTH Aachen University, Pauwelsstrasse 30, Aachen, 52074 Germany ; West Saxon University of Applied Sciences, Zwickau, Germany.

Published: May 2015

Background: The examination of joint range of motion (RoM) is part of musculo-skeletal functional diagnostics, used, for example, in occupational examinations. Various examination methodologies exist that have been optimized for occupational medical practice, which means they were reduced to the most necessary and feasible measures and examinations for efficiency and usability reasons. Because of time constraints in medical examinations in occupational settings, visual inspection is commonly used to quantify joint RoM. To support medical examiners, an inertial sensor-based measurement system (CUELA) was adapted for joint RoM examination in these settings. The objective of the present study was to evaluate the measurement tool in functional diagnostics under conditions close to clinical practice.

Methods: The joint RoM of twenty healthy subjects were examined by three physicians, who were simultaneously using the measurement tool. Physicians were blinded to the measurement results and the other physicians. Active RoM was examined on the cervical, thoracic and lumbar spine while passive RoM was examined on the shoulder, elbow, wrist, hip, and knee, resulting in a total of 40 joint examination angles. The means, standard deviations, intraclass correlation coefficients (I C C 3,k ), and Bland-Altman-Plots were calculated using MatLab for statistical analysis.

Results: Most measurement results were in accordance with expected joint RoMs. All examinations showed an acceptable repeatability. In active RoM examinations, the ICC of inter-rater reliability varied between 0.79 and 0.95. In passive RoM examination the ICC varied between 0.71 and 0.96, except examination angles at the elbow and knee extension (ICC: 0.0-0.77).

Conclusion: The reliability and objectivity of active RoM examinations were improved by the measurement tool compared with examiners. In passive RoM examinations of upper and lower extremities, the increase of objectivity by the measurements was limited for some examination angles by external factors such as the individual examiner impact on motion execution or the given joint examination conditions. Especially the elbow joint examination requires further development to achieve acceptable reliability. A modification in the examination method to reduce the examiner impact on measurement and the implementation of a more complex calibration procedure could improve the objectivity and reliability of the measurement tool in passive joint RoM examination to be applicable on nearly the whole body.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4433057PMC
http://dx.doi.org/10.1186/s12995-015-0058-5DOI Listing

Publication Analysis

Top Keywords

joint rom
16
measurement tool
16
functional diagnostics
12
rom examination
12
active rom
12
passive rom
12
joint examination
12
examination angles
12
rom examinations
12
examination
11

Similar Publications

Hip involvement is a common condition in about one-third of patients with axial spondyloarthritis (axSpA). We assessed the incidence of possible factors that predispose patients to limited flexion after total hip arthroplasty (THA) for the treatment of axSpA. We retrospectively reviewed 516 patients with axSpA (759 hips) who underwent THA.

View Article and Find Full Text PDF

Background: Calcaneal fracture malunion (CFM) commonly occurs with multiple pathologic changes and progressive pain and difficulty walking. The purpose of this study was to propose a modified 3-plane joint-preserving osteotomy for the treatment of CFM with subtalar joint incongruence, and to compare its efficacy to subtalar arthrodesis.

Methods: A retrospective comparative analysis of the data of 56 patients with CFM admitted from January 2017 to December 2022 was performed.

View Article and Find Full Text PDF

This study evaluated the validity and reliability of the TecnoBody D-Wall system in assessing joint range of motion (ROM) during overhead squat movements in healthy individuals, using Kinovea as a reference tool for data comparison. A total of 29 participants (16 males, 13 females) with a mean age of 28.41 ± 6.

View Article and Find Full Text PDF

: Measuring joint range of motion (ROM) is essential for diagnosing and treating musculoskeletal diseases. However, most clinical measurements are conducted using conventional devices, and their reliability may significantly depend on the tester. This study implemented an RGB-D (red/green/blue-depth) sensor-based artificial intelligence (AI) device to measure joint ROM and compared its reliability with that of a universal goniometer (UG).

View Article and Find Full Text PDF

Background: Knee osteoarthritis (KOA) is a prevalent degenerative joint disease. The primary pathological manifestations of KOA include articular cartilage degeneration, joint space narrowing, and osteophyte formation, leading to a spectrum of symptoms, including joint pain, stiffness, reduced mobility, diminished muscle strength, and severe disability. We aimed to utilize a meta-analysis to evaluate the efficacy of isokinetic muscle strengthening training (IMST) as a rehabilitation treatment for KOA in lowland areas.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!