Lessons from the real bench: non-BRS.

EuroIntervention

Mercy Angiography, Auckland, New Zealand.

Published: February 2016

AI Article Synopsis

  • Bench testing of bifurcation stents evaluates crucial performance factors such as crossing profile, radial strength, and flexibility, helping clarify device delivery issues.
  • Testing revealed that side branch dilatation can distort stents, leading to design improvements like adding connectors for better stability and a recommended two-step deployment strategy for optimal results.
  • Scanning electron microscopy and repetitive bend tests identify potential stent fractures and causes of balloon passage obstruction, underscoring the importance of bench testing in enhancing stent design and validating computer models for bifurcation treatments.

Article Abstract

Bench testing of stents used in bifurcations can provide information on the general properties that influence performance including crossing profile, radial strength, recoil, flexibility and radiopacity. Problems with device delivery can be clarified. Bench testing identified that side branch dilatation caused stent distortion and elucidated correction strategies. Bench testing led to a stent design change adding connectors between hoops to help overcome the clinical problem of longitudinal distortion. Testing on the bench can determine best deployment strategies and showed that a two-step post-dilatation strategy produced the best deployment with "crush" stenting. Scanning electron microscopy showed that withdrawal of a coronary guidewire trapped between a stent (or scaffold) and a mock arterial wall during a provisional side branch stenting strategy caused only mild linear polymer coating damage. Stent fracture can cause adverse clinical events and our repetitive bend test identified the stents most resistant to fracture. Causes of obstruction of the passage of a balloon over a wire through the side of a stent include damage to the catheter tip, complex cell geometry and inadvertent passage of a wire behind a strut. Bench testing plays a major role in validation of computer modelling of bifurcation treatments and flow alterations.

Download full-text PDF

Source
http://dx.doi.org/10.4244/EIJV11SVA6DOI Listing

Publication Analysis

Top Keywords

bench testing
16
side branch
8
best deployment
8
bench
6
testing
5
stent
5
lessons real
4
real bench
4
bench non-brs
4
non-brs bench
4

Similar Publications

Comprehensive Research Facility for Fusion Technology (CRAFT) is a technology development and validation platform for fusion technology in China. Neutral beam injection is one of the most important auxiliary heating and current drive methods in magnetically confined controlled fusion. Consequently, a negative ion based neutral beam injector (NNBI) testing facility with a beam energy of 400 keV is being developed in CRAFT.

View Article and Find Full Text PDF

Hemodialysis for chronic kidney disease (CKD) relies on vascular access (VA) devices, such as arteriovenous fistulas (AVF), grafts (AVG), or catheters, to maintain blood flow. Nonetheless, unpredictable progressive vascular stenosis due to neointimal formation or complete occlusion from acute thrombosis remains the primary cause of mature VA failure. Despite emergent surgical intervention efforts, the lack of a reliable early detection tool significantly reduces patient outcomes and survival rates.

View Article and Find Full Text PDF

Background: It is still unknown if the double-femoral tunnel technique (Arciero [ARC]) provides better stability as compared with the single-femoral tunnel technique (modified Larson [LAR]) in posterolateral corner reconstruction. The ideal angle of fixation of the popliteofibular strand in ARC is also unknown.

Hypotheses: The ARC provides greater external rotation (ER) stability than the LAR (hypothesis 1); there is no difference in varus rotation (VR) stability between LAR and ARC (hypothesis 2); and femoral fixation of the popliteofibular strand at 60° during the ARC leads to greater ER stability than fixation at 30° or 90° of knee flexion (hypothesis 3).

View Article and Find Full Text PDF

Purpose: Although previous studies have compared strength-training adaptations between free weights (FW) and machine-guided exercises, those studies did not use a Smith machine (SM), which most closely replicates the exercises performed with FW. Thus, the aim of the present study was to investigate the chronic effects of strength-focused, velocity-based training regimens using FW versus SM.

Methods: Thirty-seven sport-science students (14 female) were assigned, balanced by sex and relative strength, to either an FW or SM training group.

View Article and Find Full Text PDF

Advancing microfluidic point-of-care platelet function tests: opportunities and challenges from bench to market.

Front Bioeng Biotechnol

December 2024

George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States.

Platelets are critical for blood clotting, with shear-induced platelet aggregation (SIPA) playing a key role in hemostasis and the prevention of excessive bleeding. SIPA function potentially leads to life-threatening diseases such as hemorrhage and myocardial infarction, which are leading causes of death globally. Point-of-care platelet function tests (POC PFTs) are developed to assess platelet dysfunction and distinguish between normal and abnormal platelet activity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!