This article presents examination of a failed total hip replacement with surface engineered metal-on-metal (MoM) articulation. The implant was coated with a thin TiNbN film (Physical Vapor Deposition), and at retrieval 53 months after implantation the coating was abraded on weight-bearing part of the head and acetabular component. Scanning electron microscopy of bearing surfaces demonstrated multifocal crack formation, delamination of small film fragments, and formation of aggregates containing nanometer sized wear debris. We also observed coating damage in third body mechanism. Complex failure mechanism of TiNbN coating demonstrated in this study suggests insufficient bonding strength between the coating and substrate alloy and raise concerns regarding the use of such coatings in total hip arthroplasty with MoM bearing. © 2015 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 104B: 1043-1049, 2016.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jbm.b.33421 | DOI Listing |
Materials (Basel)
May 2024
Semillero Lún, Facultad de Ingeniería, Universidad de Ibagué, Ibagué 730002, Colombia.
Hardness is one of the most crucial mechanical properties, serving as a key indicator of a material's suitability for specific applications and its resistance to fracturing or deformation under operational conditions. Machine learning techniques have emerged as valuable tools for swiftly and accurately predicting material behavior. In this study, regression methods including decision trees, adaptive boosting, extreme gradient boosting, and random forest were employed to forecast Vickers hardness values based solely on scanned monochromatic images of indentation imprints, eliminating the need for diagonal measurements.
View Article and Find Full Text PDFJ Biomed Mater Res B Appl Biomater
January 2024
Department of Orthopaedics, Biomechanics and Implant Technology Research Laboratory, Rostock University Medical Center, Rostock, Germany.
To improve the wear resistance of articulating metallic joint endoprostheses, the surfaces can be coated with titanium niobium nitride (TiNbN). Under poor tribological conditions or malalignment, wear can occur on these implant surfaces in situ. This study investigated the biological response of human osteoblasts to wear particles generated from TiNbN-coated hip implants.
View Article and Find Full Text PDFMaterials (Basel)
June 2023
Facultad de Ciencias Naturales y Matemáticas, Universidad de Ibagué, Carrera 22 Calle 67, Ibagué 730002, Colombia.
This study focused on investigating the adhesion and tribological properties of niobium-doped titanium nitride (TiNbN) coatings deposited on D2 steel substrates at various substrate temperatures (Ts) under simulated cutting conditions. X-ray diffraction confirmed the presence of coatings with an FCC crystalline structure, where Nb substitutes Ti atoms in the TiN lattice. With increasing Ts, the lattice parameter decreased, and the crystallite material transitioned from flat-like to spherical shapes.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
August 2023
Laboratory of Biomechanics and Implant Research, Department of Orthopaedics, Heidelberg University Hospital, Schlierbacher Landstrasse 200a, 69118, Heidelberg, Germany.
Background: In total knee arthroplasty, femoral components with coated or oxidized surfaces are commonly used as alternative to CoCrMo in metal sensitive patients. Data on the in vivo behaviour of different coating types is, however, rare. Aim of the study was the investigation of coating stability with respect to implant and patient specific parameters.
View Article and Find Full Text PDFJ Imaging
December 2022
Professional School of Systems Engineering, Faculty of Engineering, Architecture and Urban Planning, Universidad Señor de Sipán, Chiclayo 14000, Lambayeque, Peru.
Hardness is one of the most important mechanical properties of materials, since it is used to estimate their quality and to determine their suitability for a particular application. One method of determining quality is the Vickers hardness test, in which the resistance to plastic deformation at the surface of the material is measured after applying force with an indenter. The hardness is measured from the sample image, which is a tedious, time-consuming, and prone to human error procedure.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!