Background: Hyperglycemia and hypoglycemia are associated with adverse clinical outcomes in intensive care patients. In product development studies at 4 ICUs, the safety and performance of an intravascular continuous glucose monitoring (IV-CGM) system was evaluated in 70 postsurgical patients.
Methods: The GluCath System (GluMetrics, Inc) used a quenched chemical fluorescence mechanism to optically measure blood glucose when deployed via a radial artery catheter or directly into a peripheral vein. Periodic ultrasound assessed blood flow and thrombus formation. Patient glucose levels were managed according to the standard of care and existing protocols at each site. Reference blood samples were acquired hourly and compared against prospectively calibrated sensor results.
Results: In all, 63 arterial sensors and 9 venous sensors were deployed in 70 patients. Arterial sensors did not interfere with invasive blood pressure monitoring, sampling or other aspects of patient care. A majority of venous sensors (66%) exhibited thrombus on ultrasound. In all, 89.4% (1383/1547) of arterial and 72.2% (182/252) of venous measurements met ISO15197:2003 criteria (within 20%), and 72.7% (1124/1547) of arterial and 56.3% (142/252) of venous measurements met CLSI POCT 12-A3 criteria (within 12.5%). The aggregate mean absolute relative difference (MARD) between the sensors and the reference was 9.6% for arterial and 14.2% for venous sensors.
Conclusions: The GluCath System exhibited acceptable accuracy when deployed in a radial artery for up to 48 hours in ICU patients after elective cardiac surgery. Accuracy of venous deployment was substantially lower with significant rates of intravascular thrombus observed using ultrasound.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4525661 | PMC |
http://dx.doi.org/10.1177/1932296815585872 | DOI Listing |
Anal Chim Acta
May 2025
College of Textile Science and Engineering, Jiangnan University, 1800Lihu Road, Wuxi, 214122, China. Electronic address:
Flexible sweat sensors possess the special potential for continuous non-invasive monitoring of human blood glucose. We put forward a flexible microcolumn array sensor, which is designed for health monitoring by means of detecting glucose levels in sweat and capturing physiological signals related to human movement. With the combination of silk fibroin (SF), waterborne polyurethane (PU), and multi-walled carbon nanotubes (MWCNT), this microcolumn film electrode is able to effectively function as a strain sensor benefiting from the superior mechanical performance of PU.
View Article and Find Full Text PDFBiol Pharm Bull
March 2025
Department of Pharmacy, College of Pharmacy, Kinjo Gakuin University, 2-1723 Omori, Moriyama-ku, Nagoya 463-8521, Japan.
We examined whether the glucose levels and awareness of individuals without diabetes changed after using a sensor-based intermittently scanned continuous glucose monitoring (isCGM) system in their daily lives. Japanese individuals without a diabetes diagnosis wore the isCGM system while maintaining a normal lifestyle during the baseline period. A certified diabetes educator coached them on how to improve their lifestyle based on information from sensor data, food journals, and body composition.
View Article and Find Full Text PDFJ Nutr
March 2025
Frontier Science Research Center, University of Miyazaki, Miyazaki 889-1692, Japan. Electronic address:
Background: Type 2 diabetes (T2D) is characterized by insulin resistance and defective insulin secretion. Previously, we found that rats fed soft pellets (SPs) on a 3-hour restricted schedule over 14 weeks demonstrated glucose intolerance and insulin resistance with disruption of insulin signaling.
Objective: To determine (1) the time required for an SP diet to induce insulin resistance, and (2) whether the metabolic derangements in rats fed SPs can be reversed by changing to a standard control diet.
Diabetes Metab J
March 2025
Division of Endocrinology and Metabolism, Department of Internal Medicine, St. Vincent's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea.
Backgruound: Diabetes in older adults is becoming a significant public burden to South Korea. However, a comprehensive understanding of epidemiologic trends and the detailed clinical characteristics of older adults with diabetes is lacking. Therefore, we evaluated epidemiologic trends and the metabolic and lifestyle characteristics of diabetes in Korean older adults.
View Article and Find Full Text PDFArch Gerontol Geriatr
March 2025
School of Public Health, Capital Medical University, NO. 10 Xitoutiao, Youanmenwai, Fengtai District, Beijing 100069, China. Electronic address:
Background: Insulin resistance (IR) is strongly related to non-alcoholic fatty liver disease (NAFLD). Triglyceride-glucose (TyG) index serves as a novel substitute indicator for IR. However, research on the effect of TyG index on NAFLD remains sparse.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!