Background: Reconstruction of the coracoclavicular ligament functions to restore anatomic alignment of the clavicle and may improve biomechanical function and clinical outcomes. Improper placement of the coracoclavicular tunnel may inherently weaken the coracoid. The purpose of this study was to evaluate the feasibility and accuracy of navigated image-free placement of K-wires for coracoclavicular tunnel position in comparison to conventional drill guide-based placement.

Materials And Methods: Eight human shoulder specimens were assigned for conventional technique with a coracoclavicular guide device (group CP) and the paired contralateral side for the navigated procedure (group NP) with an optoelectronic system with a fluoro-free software module. First-pass accuracy (%) and the K-wire trajectory (lateral-center orientation (LC), center-center (CC) orientation and medial-center orientation (MC) were measured.

Results: In all navigated K-wires a 100 % first-pass accuracy was observed. In three of the eight (37.5 %) specimens of the drill guide-based group, drilling had to be repeated. One of them had to be repeated twice, resulting in eight versus twelve drillings for the navigated versus conventional group, respectively (p = 0.021). K-wire trajectory showed an MC orientation in most of the specimen (n = 9, group NP 4, group CP 5).

Conclusions: Image-free navigated coracoclavicular drilling for the repair of acromioclavicular joint dislocation has higher first-pass accuracy in comparison to conventional drill guide-based placement and, therefore, may enable a precise anatomic position of the drill holes and reduce the risk of an iatrogenic coracoid fracture.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00402-015-2243-5DOI Listing

Publication Analysis

Top Keywords

drill guide-based
12
first-pass accuracy
12
image-free navigated
8
navigated coracoclavicular
8
coracoclavicular drilling
8
drilling repair
8
repair acromioclavicular
8
acromioclavicular joint
8
joint dislocation
8
coracoclavicular tunnel
8

Similar Publications

Background: Accurate implant placement in the bone is key to successful implant treatment. Once inserted, it can be difficult to correct the orientation of the implant axis, especially of a one-piece implant. Prosthetic-driven digital implant planning in combination with fully guided implant surgery can offer additional safety in such cases.

View Article and Find Full Text PDF

Purpose: Computer-aided surgery under navigation system guidance is widely applied in dental implant procedures. However, the accuracy of drilling with such navigation systems has not been comparatively evaluated alongside those of laboratory guide-based and freehand drilling. Therefore, this study aimed to compare the accuracies of these three drilling systems.

View Article and Find Full Text PDF

Background: The design and application of assisted instrumentation for internal fixation of femoral neck fractures with multiple screws are still evolving. A novel guide based on a femoral neck section was designed to improve the accuracy of screw placement, and its efficacy was evaluated.

Methods: A guide based on a femoral neck section was designed for assisted fixation of femoral neck fractures with multiple screws.

View Article and Find Full Text PDF

Arthroscopically guided navigation for repair of acromioclavicular joint dislocations: a safe technique with reduced intraoperative radiation exposure.

Patient Saf Surg

December 2015

Department of Orthopaedics, Trauma and Plastic Surgery, University of Leipzig, Liebigstr. 20, 04103 Leipzig, Germany ; Devision of Arthroscopy, Joint Surgery and Sport Injuries, University of Leipzig, Liebigstr. 20, 04103 Leipzig, Germany.

Background: Accuracy evaluation of navigated image free placement of double cortical fixation buttons for coracoclavicular tunnel position in comparison to conventional drill guide based placement.

Methods: Twenty-six patients with acute acromioclavicular joint instability were included in this non-randomized cohort study. All patients were treated with a Double- TightRope technique.

View Article and Find Full Text PDF

Background: Reconstruction of the coracoclavicular ligament functions to restore anatomic alignment of the clavicle and may improve biomechanical function and clinical outcomes. Improper placement of the coracoclavicular tunnel may inherently weaken the coracoid. The purpose of this study was to evaluate the feasibility and accuracy of navigated image-free placement of K-wires for coracoclavicular tunnel position in comparison to conventional drill guide-based placement.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!