Background/aims: A barbed puncture needle with a side hole was recently developed to improve sample quality and quantity in endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA). In this study, we retrospectively assessed the usefulness of this puncture needle.

Methods: Factors affecting diagnostic yield, safety, and diagnostic accuracy were investigated in 76 patients who consecutively underwent EUS-FNA for neoplastic lesions at our hospital between January and December 2013.

Results: The procedure was successful in all cases; the rates of sample collection and determination of the correct diagnosis were 92.1% and 89.5%, respectively. The mean number of needle passes required for diagnosis was 1.1. Complications included mild intraluminal bleeding in two patients (2.6%). Multivariate analysis revealed that lesion size (≤20 mm) was significantly associated with a decreased chance of determining the correct diagnosis.

Conclusions: Core biopsy needles with a side hole are safe and provide a satisfactory diagnostic yield. However, the side hole may potentially reduce the rate of making the correct diagnosis in small lesions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4694734PMC
http://dx.doi.org/10.5009/gnl14249DOI Listing

Publication Analysis

Top Keywords

side hole
16
diagnostic yield
12
factors diagnostic
8
core biopsy
8
biopsy needles
8
needles side
8
endoscopic ultrasound-guided
8
ultrasound-guided fine-needle
8
fine-needle aspiration
8
correct diagnosis
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!