Objective: Accuracy of continuous glucose monitoring (CGM) devices in hypoglycemia has been a widely reported shortcoming of this technology. We report the accuracy in hypoglycemia of a new version of the Dexcom (San Diego, CA) G4 Platinum CGM system (software 505) and present results regarding the optimum setting of CGM hypoglycemic alerts.
Materials And Methods: CGM values were compared with YSI analyzer (YSI Life Sciences, Yellow Springs, OH) measurements every 15 min. We reviewed the accuracy of the CGM system in the hypoglycemic range using standard metrics. We analyzed the time required for the CGM system to detect biochemical hypoglycemia (70 mg/dL) compared with the YSI with alert settings at 70 mg/dL and 80 mg/dL. We also analyzed the time between the YSI value crossing 55 mg/dL, defined as the threshold for cognitive impairment due to hypoglycemia, and when the CGM system alerted for hypoglycemia.
Results: The mean absolute difference for a glucose level of less than 70 mg/dL was 6 mg/dL. Ninety-six percent of CGM values were within 20 mg/dL of the YSI values between 40 and 80 mg/dL. When the CGM hypoglycemic alert was set at 80 mg/dL, the device provided an alert for biochemical hypoglycemia within 10 min in 95% of instances and at least a 10-min advance warning before the cognitive impairment threshold in 91% of instances in the study.
Conclusions: Use of an 80 mg/dL threshold setting for hypoglycemic alerts on the G4 Platinum (software 505) may provide patients with timely warning of hypoglycemia before the onset of cognitive impairment, enabling them to treat themselves for hypoglycemia with fast-acting carbohydrates and prevent neuroglycopenia associated with very low glucose levels.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1089/dia.2014.0415 | DOI Listing |
Diabet Med
January 2025
Diabetes and Endocrine Centre, Sheffield Teaching Hospitals, Sheffield, UK.
Aims: This study assessed real-world glycaemic outcomes associated with the use of Dexcom ONE in adults with suboptimally controlled diabetes.
Methods: In this single-site prospective study, adults with type 1 (T1D) or type 2 diabetes (T2D) taking two or more insulin injections per day initiated Dexcom ONE CGM use and attended follow-up data collection visits after 3 and 6 months. During the study, participants received usual diabetes care.
Diabetes Technol Ther
January 2025
Senseonics, Incorporated, Germantown, Maryland, USA.
The implanted Eversense Continuous Glucose Monitoring (CGM) System transitioned from 90- to 180- to 365-day durations marketed today. This report summarizes the 365-day clinical study. ENHANCE was a prospective, multicenter study evaluating the accuracy and safety of the Eversense 365 CGM system through 1 year in adults with diabetes.
View Article and Find Full Text PDFDiabetes Obes Metab
January 2025
Department of Endocrinology and Metabolism, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai Clinical Center for Diabetes, Shanghai Diabetes Institute, Shanghai Key Laboratory of Diabetes Mellitus, Shanghai, China.
Aims: Currently, there is a lack of evidence regarding time in tight range (TITR) and long-term adverse outcomes. We aimed to investigate the association between TITR and the risk of all-cause and cardiovascular mortality among patients with type 2 diabetes.
Materials And Methods: A total of 6061 patients with type 2 diabetes were prospectively recruited in a single centre.
JCEM Case Rep
February 2025
Division of Endocrinology, Diabetes & Metabolic Diseases, Thomas Jefferson University Hospital, Philadelphia, PA 19107, USA.
Insulinomas are rare neuroendocrine neoplasms and causes of hypoglycemia. They present with neuroglycopenic symptoms, including confusion and seizures. Suspected diagnosis must be confirmed through bloodwork and imaging.
View Article and Find Full Text PDFBackground: The use of automated insulin delivery (AID) devices is now widespread in the management of type 1 diabetes (T1D), being used for younger and older children, adolescents and adults. The integration of insulin pumps with continuous glucose monitors (CGM) and smart management software in AID systems has significantly improved glycemic management compared to the separate application of each diabetes technology. The efficacy of AID systems has been demonstrated in randomized controlled trials (RCTs) but it is their application in real-world studies that fully demonstrates their impact for people with T1D.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!