Statement Of Problem: Controlling tooth reduction for porcelain laminate veneers (PLVs) in fractions of millimeters is challenging.
Purpose: The purpose of this study was to assess an automated robotic tooth preparation system for PLVs for accuracy and precision compared with conventional freehand tooth preparation.
Material And Methods: Twenty maxillary central incisor tooth models were divided into 2 groups. Ten were assigned to a veneer preparation with a robotic arm according to preoperative preparation design-specific guidelines (experimental group). Ten were assigned to conventional tooth preparation by a clinician (control group). Initially, all tooth models were scanned with a 3- dimensional (3D) laser scanner, and a tooth preparation for PLVs was designed on a 3D image. Each tooth model was attached to a typodont. For the experimental group, an electric high-speed handpiece with a 0.9-mm-diameter round diamond rotary cutting instrument was mounted on the robotic arm. The teeth were prepared automatically according to the designed image. For the control group, several diamond rotary cutting instruments were used to prepare the tooth models according to preoperative preparation design guidelines. All prepared tooth models were scanned. The preoperative preparation design image and scanned postoperative preparation images were superimposed. The dimensional difference between those 2 images was measured on the facial aspect, finish line, and incisal edge. Differences between the experimental and the control groups from the 3D design image were computed. Accuracy and precision were compared for all sites and separately for each tooth surface (facial, finish line, incisal). Statistical analyses were conducted with a permutation test for accuracy and with a modified robust Brown-Forsythe Levene-type test for precision (α=.05).
Results: For accuracy for all sites, the mean absolute deviation was 0.112 mm in the control group and 0.133 mm in the experimental group. No significant difference was found between the 2 (P=.15). For precision of all sites, the standard deviation was 0.141 mm in the control group and 0.185 mm in the experimental group. The standard deviation in the control group was significantly lower (P=.030). In terms of accuracy for the finish line, the control group was significantly less accurate (P=.038). For precision, the standard deviation in the control group was significantly higher at the finish line (P=.034).
Conclusions: For the data from all sites, the experimental procedure was able to prepare the tooth model as accurately as the control, and the control procedure was able to prepare the tooth model with better precision. The experimental group showed better accuracy and precision at the finish line.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.prosdent.2015.02.021 | DOI Listing |
J Ultrasound
January 2025
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, University Hospital and Health Services of Trieste, ASUGI, University of Trieste, Strada di Fiume, 447, 34149, Trieste, Italy.
Introduction: Post-stroke cognitive impairment (PSCI) and dementia affect short- and long-term outcome after stroke and can persist even after recover from a physical handicap. The process underlying PSCI is not yet fully understood. Transcranial Doppler ultrasound (TCD) is a feasible method to investigate cerebrovascular aging or dementia, through the pulsatility index (PI), the cerebrovascular reactivity (e.
View Article and Find Full Text PDFEndocrine
January 2025
Department of Endocrinology and Metabolic Diseases, Manisa Celal Bayar University Hospital, Manisa, Turkey.
Purpose: Our study evaluated skeletal muscle mass, function and quality among mild autonomous cortisol secretion (MACS) patients and non-functioning adrenal incidentaloma (NFAI) patients in comparison with the control group without adrenal mass.
Methods: 63 NFAI (49 female, 14 male) and 31 MACS (24 female, 7 male) patients were included in the study. As the control group, 44 patients (31 women, 13 men) who were known to have no radiological adrenal pathology on computed tomography or magnetic resonance imaging performed for other reasons were selected.
Cardiovasc Toxicol
January 2025
Department of Morphological Sciences, State University of Maringa, Maringa, Parana, Brazil.
5-Fluorouracil (5-FU) is a chemotherapeutic that is used to treat solid tumors. However, 5-FU is associated with several side effects, including cardiotoxicity. Considering the importance of the intrinsic cardiac nervous system (ICNS) for the heart and that little is known about effects of 5-FU on this nervous system plexus, the purpose of the present study was to evaluate effects 5-FU at a low dose on the ICNS and oxidative and inflammatory effects in the heart in Wistar rats.
View Article and Find Full Text PDFJ Perianesth Nurs
January 2025
Medical Faculty, Department of Anesthesia and Reanimation, Giresun Üniversitesi, Giresun, Türkiye. Electronic address:
Purpose: The aim in the present study was to evaluate the effects of chewing postoperative xylitol gum on gastrointestinal functional recovery after laparoscopic cholecystectomy.
Design: A three-arm randomized controlled trial.
Methods: After a baseline assessment, participants were randomly assigned in three groups.
Dis Esophagus
January 2025
Department of Digestive and Oncological Surgery, Claude Huriez Hospital, Chu Lille, Lille, France.
Background: Malnutrition is common with esophagogastric cancers and is associated with negative outcomes. We aimed to evaluate if immunonutrition during neoadjuvant treatment improves patient's health-related quality of life (HRQOL) and reduces postoperative morbidity and toxicities during neoadjuvant treatment.
Methods: A multicenter double-blind randomized controlled trial (RCT) was undertaken.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!