Purpose: The aim was to compare the corneal curvature and power measured with a corneal topographer, Scheimpflug camera, optical biometer, and Javal keratometer.
Materials And Methods: A total of 76 myopic individuals who were candidates for photorefractive keratectomy were selected in a cross-sectional study. Manual keratometry (Javal Schiotz type; Haag-Streit AG, Koeniz, Switzerland), automated keratometry (IOL Master version 3.02, Carl Zeiss Meditec, Jena, Germany), topography (TMS4, Tomey, Erlangen, Germany), and Pentacam HR (Oculus, Wetzlar, Germany) were performed for all participants. The 95% limits of agreement (LOAs) were reported to evaluate the agreement between devices.
Results: The mean corneal power measurements were 44.3 ± 1.59, 44.25 ± 1.59, 43.68 ± 1.44, and 44.31 ± 1.61 D with a Javal keratometer, TMS4-topographer, the Pentacam and IOL Master respectively. Only the IOL Master showed no significant difference with Javal keratometer in measuring the corneal power (P = 0.965). The correlations of the Javal keratometer with TMS4-topography, Pentacam, and IOL Master was 0.991. 0.982, and 0.993 respectively. The 95% LOAs of the Javal keratometer with TMS4-topography, Pentacam, and IOL Master were - 0.361 to 0.49, -0.01 to 1.14, and - 0.36 to 0.36 D, respectively.
Conclusion: Although the correlation of Pentacam, TMS4-topography, IOL Master, and Javal keratometer in measuring keratometry was high, only the IOL Master showed no significant difference with the Javal keratometer. The IOL Master had the best agreement with Javal keratometry.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4411623 | PMC |
http://dx.doi.org/10.4103/0974-9233.151884 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!