Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Device longevity in cardiac resynchronization therapy (CRT) is affected by the pacing capture threshold (PCT) and programmed pacing amplitude of the left ventricular (LV) pacing lead. The aims of this study were to evaluate the stability of LV pacing thresholds in a nationwide sample of CRT defibrillator recipients and to determine potential longevity improvements associated with a decrease in the LV safety margin while maintaining effective delivery of CRT. CRT defibrillator patients in the Medtronic CareLink database were eligible for inclusion. LV PCT stability was evaluated using ≥2 measurements over a 14-day period. Separately, a random sample of 7,250 patients with programmed right atrial and right ventricular amplitudes ≤2.5 V, LV thresholds ≤ 2.5 V, and LV pacing ≥90% were evaluated to estimate theoretical battery longevity improvement using LV safety margins of 0.5 and 1.5 V. Threshold stability analysis in 43,256 patients demonstrated LV PCT stability of <0.5 V in 77% of patients and <1 V in 95%. Device longevity analysis showed that the use of a 0.5-V safety margin increased average battery longevity by 0.62 years (95% confidence interval 0.61 to 0.63) compared with a safety margin of 1.5 V. Patients with LV PCTs >1 V had the greatest increases in battery life (mean increase 0.86 years, 95% confidence interval 0.85 to 0.87). In conclusion, nearly all CRT defibrillator patients had LV PCT stability <1.0 V. Decreasing the LV safety margin from 1.5 to 0.5 V provided consistent delivery of CRT for most patients and significantly improved battery longevity.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4480976 | PMC |
http://dx.doi.org/10.1016/j.amjcard.2015.03.047 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!